ABSTRACT: Objectives
To access the performance of 12 serological tests for COVID-19 diagnosis.Methods
We conducted a blind evaluation of six lateral-flow immunoassays (LFIAs) and six enzyme-linked immunosorbent assays (ELISAs) commercially available in Brazil for detecting anti-SARS-CoV-2 antibodies.Results
Considering patients with seven or more days of symptoms, the sensitivity ranged from 59.5% to 83.1% for LFIAs and from 50.7% to 92.6% for ELISAs. For both methods, the sensitivity increased with clinical severity and days of symptoms. The agreement among LFIAs performed with digital blood and serum was moderate. Specificity was, in general, higher for LFIAs than for ELISAs. Infectious diseases prevalent in the tropics, such as HIV, leishmaniasis, arboviruses, and malaria, represent conditions with the potential to cause false-positive results with these tests, which significantly compromises their specificity.Conclusion
The performance of immunoassays was only moderate, affected by the duration and clinical severity of the disease. Absence of discriminatory power between IgM/IgA and IgG has also been demonstrated, which prevents the use of acute-phase antibodies for decisions on social isolation.