Validation of thoracic aortic dimensions on ECG-triggered SSFP as alternative to contrast-enhanced MRA.


ABSTRACT: OBJECTIVES:Assessment of thoracic aortic dimensions with non-ECG-triggered contrast-enhanced magnetic resonance angiography (CE-MRA) is accompanied with motion artefacts and requires gadolinium. To avoid both motion artefacts and gadolinium administration, we evaluated the similarity and reproducibility of dimensions measured on ECG-triggered, balanced steady-state free precession (SSFP) MRA as alternative to CE-MRA. METHODS:All patients, with varying medical conditions, referred for thoracic aortic examination between September 2016 and March 2018, who underwent non-ECG-triggered CE-MRA and SSFP-MRA (1.5 T) were retrospectively included (n = 30). Aortic dimensions were measured after double-oblique multiplanar reconstruction by two observers at nine landmarks predefined by literature guidelines. Image quality was scored at the sinus of Valsalva, mid-ascending aorta and mid-descending aorta by semi-automatically assessing the vessel sharpness. RESULTS:Aortic dimensions showed high agreement between non-ECG-triggered CE-MRA and SSFP-MRA (r = 0.99, p < 0.05) without overestimation or underestimation of aortic dimensions in SSFP-MRA (mean difference, 0.1 mm; limits of agreement, -?1.9 mm and 1.9 mm). Intra- and inter-observer variabilities were significantly smaller with SSFP-MRA for the sinus of Valsalva and sinotubular junction. Image quality of the sinus of Valsalva was significantly better with SSFP-MRA, as fewer images were of impaired quality (3/30) than in CE-MRA (21/30). Reproducibility of dimensions was significantly better in images scored as good quality compared to impaired quality in both sequences. CONCLUSIONS:Thoracic aortic dimensions measured on SSFP-MRA and non-ECG-triggered CE-MRA were similar. As expected, SSFP-MRA showed better reproducibility close to the aortic root because of lesser motion artefacts, making it a feasible non-contrast imaging alternative. KEY POINTS:• SSFP-MRA provides similar dimensions as non-ECG-triggered CE-MRA. • Intra- and inter-observer reproducibilities improve for the sinus of Valsalva and sinotubular junction with SSFP-MRA. • ECG-triggered SSFP-MRA shows better image quality for landmarks close to the aortic root in the absence of cardiac motion.

SUBMITTER: Snel GJH 

PROVIDER: S-EPMC7554008 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of thoracic aortic dimensions on ECG-triggered SSFP as alternative to contrast-enhanced MRA.

Snel G J H GJH   Hernandez L M LM   Slart R H J A RHJA   Nguyen C T CT   Sosnovik D E DE   van Deursen V M VM   Dierckx R A J O RAJO   Velthuis B K BK   Borra R J H RJH   Prakken N H J NHJ  

European radiology 20200607 11


<h4>Objectives</h4>Assessment of thoracic aortic dimensions with non-ECG-triggered contrast-enhanced magnetic resonance angiography (CE-MRA) is accompanied with motion artefacts and requires gadolinium. To avoid both motion artefacts and gadolinium administration, we evaluated the similarity and reproducibility of dimensions measured on ECG-triggered, balanced steady-state free precession (SSFP) MRA as alternative to CE-MRA.<h4>Methods</h4>All patients, with varying medical conditions, referred  ...[more]

Similar Datasets

| S-EPMC4779553 | biostudies-literature
| S-EPMC4689113 | biostudies-literature
| S-EPMC6728222 | biostudies-literature
| S-EPMC4452776 | biostudies-literature
| S-EPMC5932860 | biostudies-literature
| S-EPMC2676130 | biostudies-literature
| S-EPMC10795846 | biostudies-literature
| S-EPMC7590002 | biostudies-literature
| S-EPMC8118944 | biostudies-literature
| S-EPMC7263981 | biostudies-literature