Making a Molehill out of a Mountain: A Case of Transesophageal Echocardiographic Probe Malfunction


ABSTRACT: Graphical abstract Highlights • Probe malfunction during TEE is an underrecognized problem.• MR severity assessment: host (e.g., BP), technical (e.g., gain), probe malfunction.• Clues for malfunction 1: disparity between TTE and TEE; CW and color Doppler.• Clues for malfunction 2: unexplained poor image quality; need for high gain.• Transducer temperature can affect image quality.