Four-dimensional impedance manometry derived from esophageal high-resolution impedance-manometry studies: a novel analysis paradigm.
Ontology highlight
ABSTRACT: Background:This study aimed to introduce a novel analysis paradigm, referred to as 4-dimensional (4D) manometry based on biophysical analysis; 4D manometry enables the visualization of luminal geometry of the esophagus and esophagogastric junction (EGJ) using high-resolution-impedance-manometry (HRIM) data. Methods:HRIM studies from two asymptomatic controls and one type-I achalasia patient were analyzed. Concomitant fluoroscopy images from one control subject were used to validate the calculated temporal-spatial luminal radius and time-history of intraluminal bolus volume and movement. EGJ analysis computed diameter threshold for emptying, emptying time, flow rate, and distensibility index (DI), which were compared with bolus flow time (BFT) analysis. Results:For normal control, calculated volumes for 5?ml swallows were 4.1?ml-6.7?ml; for 30?ml swallows 21.3?ml-21.8?ml. With type-I achalasia, >4?ml of intraesophageal bolus residual was present both pre- and post-swallow. The four phases of bolus transit were clearly illustrated on the time-history of bolus movement, correlating well with the fluoroscopic images. In the control subjects, the EGJ diameter threshold for emptying was 8?mm for 5?ml swallows and 10?mm for 30?ml swallows; emptying time was 1.2-2.2?s for 5?ml swallows (BFT was 0.3-3?s) and 3.25-3.75?s for 30?ml swallows; DI was 2.4-3.4?mm2/mmHg for 5?ml swallows and 4.2-4.6?mm2/mmHg for 30?ml swallows. Conclusions:The 4D manometry system facilitates a comprehensive characterization of dynamic esophageal bolus transit with concurrent luminal morphology and pressure from conventional HRIM measurements. Calculations of flow rate and wall distensibility provide novel measures of EGJ functionality.
SUBMITTER: Kou W
PROVIDER: S-EPMC7592175 | biostudies-literature | 2020
REPOSITORIES: biostudies-literature
ACCESS DATA