Development and Greenness Evaluation of Spectrofluorometric Methods for Flibanserin Determination in Dosage Form and Human Urine Samples.


ABSTRACT: Green, economic and sensitive two spectrofluorometric methods were developed for the quantitation of flibanserin (FB) in different matrices, which are based on FB native fluorescence properties. The first technique depends on measuring the relative fluorescence intensity of FB directly at emission and excitation wavelengths(λem/λex) (371 nm/247 nm), while the second technique is a first derivative (D1) spectrofluorometric method, which depends on measuring the peak amplitudes at 351 nm. Linear regressions were observed in the range of 0.1-1.5 μg/mL for both methods. Moreover, both methods were efficiently extended to analyze FB in human urine, indicating the ultra-sensitivity of the methods, and linear regression was found within a range 0.05-0.7 μg/mL for both methods. Excellent selectivi

SUBMITTER: Ahmed R 