Sensitive and Simultaneous Detection of SARS-CoV-2-Specific IgM/IgG Using Lateral Flow Immunoassay Based on Dual-Mode Quantum Dot Nanobeads.


ABSTRACT: A rapid and accurate method for detection of virus (SARS-CoV-2)-specific antibodies is important to contain the 2019 coronavirus disease (COVID-19) outbreak, which is still urgently needed. Here, we develop a colorimetric-fluorescent dual-mode lateral flow immunoassay (LFIA) biosensor for rapid, sensitive, and simultaneous detection of SARS-CoV-2-specific IgM and IgG in human serum using spike (S) protein-conjugated SiO2@Au@QD nanobeads (NBs) as labels. The assay only needs 1 ?L of the serum sample, can be completed within 15 min, and is 100 times more sensitive than the colloidal gold-based LFIA. Two detection modes of our biosensor are available: the colorimetric mode for rapid screening of the patients with suspected SARS-CoV-2 infection without any special instrument and the fluorescent mode for sensitive and quantitative analyses to determine the concentrations of specific IgM/IgG in human serum and detect the infection early and precisely. We validated the proposed method using 16 positive serum samples from patients with COVID-19 and 41 negative samples from patients with other viral respiratory infections. The results demonstrated that combined detection of virus-specific IgM and IgG via SiO2@Au@QD LFIA can identify 100% of patients with SARS-CoV-2 infection with 100% specificity.

SUBMITTER: Wang C 

PROVIDER: S-EPMC7687295 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Sensitive and Simultaneous Detection of SARS-CoV-2-Specific IgM/IgG Using Lateral Flow Immunoassay Based on Dual-Mode Quantum Dot Nanobeads.

Wang Chongwen C   Yang Xingsheng X   Gu Bing B   Liu Haifeng H   Zhou Zihui Z   Shi Luoluo L   Cheng Xiaodan X   Wang Shengqi S  

Analytical chemistry 20201118 23


A rapid and accurate method for detection of virus (SARS-CoV-2)-specific antibodies is important to contain the 2019 coronavirus disease (COVID-19) outbreak, which is still urgently needed. Here, we develop a colorimetric-fluorescent dual-mode lateral flow immunoassay (LFIA) biosensor for rapid, sensitive, and simultaneous detection of SARS-CoV-2-specific IgM and IgG in human serum using spike (S) protein-conjugated SiO<sub>2</sub>@Au@QD nanobeads (NBs) as labels. The assay only needs 1 μL of th  ...[more]

Similar Datasets

| S-EPMC3915093 | biostudies-literature
| S-EPMC8413105 | biostudies-literature
| S-EPMC3231422 | biostudies-literature
| S-EPMC9048750 | biostudies-literature
| S-EPMC10721894 | biostudies-literature
| S-EPMC9040449 | biostudies-literature
| S-EPMC9737793 | biostudies-literature
| S-EPMC8746978 | biostudies-literature
| S-EPMC7673228 | biostudies-literature
| S-EPMC7495760 | biostudies-literature