Unknown

Dataset Information

0

Adaptive slice-specific z-shimming for 2D spoiled gradient-echo sequences.


ABSTRACT: Purpose: To reduce the misbalance between compensation gradients and macroscopic field gradients, we introduce an adaptive slice-specific z-shimming approach for 2D spoiled multi-echo gradient-echoe sequences in combination with modeling of the signal decay.

Methods: Macroscopic field gradients were estimated for each slice from a fast prescan (15 seconds) and then used to calculate slice-specific compensation moments along the echo train. The coverage of the compensated field gradients was increased by applying three positive and three negative moments. With a forward model, which considered the effect of the slice profile, the z-shim moment, and the field gradient, R2? maps were estimated. The method was evaluated in phantom and in vivo measurements at 3 T and compared with a spoiled multi-echo gradient-echo and a global z-shimming approach without slice-specific compensation.

Results: The proposed method yielded higher SNR in R2? maps due to a broader range of compensated macroscopic field gradients compared with global z-shimming. In global white matter, the mean interquartile range, proxy for SNR, could be decreased to 3.06 s-1 with the proposed approach, compared with 3.37 s-1 for global z-shimming and 3.52 s-1 for uncompensated multi-echo gradient-echo.

Conclusion: Adaptive slice-specific compensation gradients between echoes substantially improved the SNR of R2? maps, and the signal could also be rephased in anatomical areas, where it has already been completely dephased.

SUBMITTER: Soellradl M 

PROVIDER: S-EPMC7693070 | biostudies-literature | 2021 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Adaptive slice-specific z-shimming for 2D spoiled gradient-echo sequences.

Soellradl Martin M   Strasser Johannes J   Lesch Andreas A   Stollberger Rudolf R   Ropele Stefan S   Langkammer Christian C  

Magnetic resonance in medicine 20200910 2


<h4>Purpose</h4>To reduce the misbalance between compensation gradients and macroscopic field gradients, we introduce an adaptive slice-specific z-shimming approach for 2D spoiled multi-echo gradient-echoe sequences in combination with modeling of the signal decay.<h4>Methods</h4>Macroscopic field gradients were estimated for each slice from a fast prescan (15 seconds) and then used to calculate slice-specific compensation moments along the echo train. The coverage of the compensated field gradi  ...[more]

Similar Datasets

| S-EPMC7216950 | biostudies-literature
| S-EPMC6636815 | biostudies-literature
| S-EPMC10952755 | biostudies-literature
| S-EPMC5599252 | biostudies-literature
| S-EPMC10102749 | biostudies-literature
| S-EPMC9845189 | biostudies-literature
| S-EPMC8244023 | biostudies-literature
| S-EPMC8028796 | biostudies-literature
| S-EPMC8338872 | biostudies-literature
| S-EPMC4021580 | biostudies-literature