CBCT image quality QA: Establishing a quantitative program.


ABSTRACT: PURPOSE:Routine quality assurance (QA) of cone-beam computed tomography (CBCT) scans used for image-guided radiotherapy is prescribed by the American Association of Physicists in Medicine Task Group (TG)-142 report. For CBCT image quality, TG-142 recommends using clinically established baseline values as QA tolerances. This work examined how image quality parameters vary both across machines of the same model and across different CBCT techniques. Additionally, this work investigated how image quality values are affected by imager recalibration and repeated exposures during routine QA. METHODS:Cone-beam computed tomography scans of the Catphan 604 phantom were taken on four TrueBeam® and one Edge™ linear accelerator using four manufacturer-provided techniques. TG-142 image quality parameters were calculated for each CBCT scan using SunCHECK Machine™. The variability of each parameter with machine and technique was evaluated using a two-way ANOVA test on a dataset consisting of 200 CBCT scans. The impact of imager calibration on image quality parameters was examined for a subset of three machines using an unpaired Student's t-test. The effect of artifacts appearing on CBCTs taken in rapid succession was characterized and an approach to reduce their appearance was evaluated. Additionally, a set of baselines and tolerances for all image quality metrics was presented. RESULTS:All imaging parameters except geometric distortion varied with technique (P < 0.05) and all imaging parameters except slice thickness varied with machine (P < 0.05). Imager calibration can change the expected value of all imaging parameters, though it does not consistently do so. While changes are statistically significant, they may not be clinically significant. Finally, rapid acquisition of CBCT scans can introduce image artifacts that degrade CBCT uniformity. CONCLUSIONS:This work characterized the variability of acquired CBCT data across machines and CBCT techniques along with the impact of imager calibration and rapid CBCT acquisition on image quality.

SUBMITTER: Taneja S 

PROVIDER: S-EPMC7701111 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

CBCT image quality QA: Establishing a quantitative program.

Taneja Sameer S   Barbee David L DL   Rea Anthony J AJ   Malin Martha M  

Journal of applied clinical medical physics 20201019 11


<h4>Purpose</h4>Routine quality assurance (QA) of cone-beam computed tomography (CBCT) scans used for image-guided radiotherapy is prescribed by the American Association of Physicists in Medicine Task Group (TG)-142 report. For CBCT image quality, TG-142 recommends using clinically established baseline values as QA tolerances. This work examined how image quality parameters vary both across machines of the same model and across different CBCT techniques. Additionally, this work investigated how  ...[more]

Similar Datasets

| S-EPMC5718664 | biostudies-literature
| S-EPMC4846179 | biostudies-literature
| S-EPMC5690923 | biostudies-literature
| S-EPMC5723455 | biostudies-literature
| S-EPMC4836729 | biostudies-literature
| S-EPMC5856621 | biostudies-literature
| S-EPMC10715786 | biostudies-literature
| S-EPMC8293763 | biostudies-literature
| S-EPMC4078205 | biostudies-other
| S-EPMC4108445 | biostudies-literature