Unknown

Dataset Information

0

Cone-Beam CT Image Guidance With and Without Electromagnetic Navigation Bronchoscopy for Biopsy of Peripheral Pulmonary Lesions.


ABSTRACT:

Background

Bronchoscopic diagnosis of small peripheral lung lesions suspected of lung cancer remains a challenge. A successful endobronchial diagnosis comprises navigation, confirmation, and tissue acquisition. In all steps, 3-dimensional information is essential. Cone-beam computed tomography (CBCT) imaging can provide computed tomography information and 3-dimensional augmented fluoroscopy imaging. We assessed whether CBCT imaging can improve navigation and diagnosis of peripheral lesions by 2 clinical workflows with a cross-over design: (1) a primary CBCT and radial endobronchial ultrasound mini probe imaging-based approach and (2) a primary electromagnetic navigation (EMN) and radial endobronchial ultrasound mini probe imaging-based approach.

Methods

All patients with a peripheral lung lesion biopsy indication were eligible for study inclusion and randomly assigned to study arms. Commercially available equipment was used. The main study goals were to assess CBCT-confirmed navigation success and diagnostic accuracy. Surgery or unambiguous clinical follow-up served as the gold standard.

Results

Eighty-seven patients with 107 lesions were included. Lesion mean longest axis size in the CBCT arm was 16.6 mm (n=47) and 14.2 mm in the EMN arm (n=40). The primary CBCT approach and primary EMN approach had 76.3% and 52.2% navigation success, respectively. Addition of EMN to the CBCT approach increased navigation success to 89.9%. Addition of CBCT imaging to the EMN approach significantly increased navigation success to 87.5% per lesion. The overall diagnostic accuracy per patient was significantly lower than the navigation success, being 72.4%.

Conclusion

CBCT imaging is a valuable addition to navigation bronchoscopy. Although overall navigation success was high, the diagnostic accuracy remains to be improved. Future research should focus on improving the tissue acquisition methodology.

SUBMITTER: Verhoeven RLJ 

PROVIDER: S-EPMC7742216 | biostudies-literature | 2021 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Cone-Beam CT Image Guidance With and Without Electromagnetic Navigation Bronchoscopy for Biopsy of Peripheral Pulmonary Lesions.

Verhoeven Roel L J RLJ   Fütterer Jurgen J JJ   Hoefsloot Wouter W   van der Heijden Erik H F M EHFM  

Journal of bronchology & interventional pulmonology 20210101 1


<h4>Background</h4>Bronchoscopic diagnosis of small peripheral lung lesions suspected of lung cancer remains a challenge. A successful endobronchial diagnosis comprises navigation, confirmation, and tissue acquisition. In all steps, 3-dimensional information is essential. Cone-beam computed tomography (CBCT) imaging can provide computed tomography information and 3-dimensional augmented fluoroscopy imaging. We assessed whether CBCT imaging can improve navigation and diagnosis of peripheral lesio  ...[more]

Similar Datasets

| S-EPMC6166698 | biostudies-literature
| S-EPMC10224575 | biostudies-literature
| S-EPMC7869923 | biostudies-literature
| S-EPMC10984630 | biostudies-literature
| S-EPMC7330771 | biostudies-literature
| S-EPMC11494579 | biostudies-literature
| S-EPMC8460082 | biostudies-literature
| S-EPMC7797816 | biostudies-literature
| S-EPMC7742212 | biostudies-literature
| S-EPMC11472628 | biostudies-literature