Diagnostic accuracy of magnetic resonance imaging for the detection of pulmonary nodules simulated in a dedicated porcine chest phantom.


ABSTRACT:

Objective

CT serves as gold standard for the evaluation of pulmonary nodules. However, CT exposes patients to ionizing radiation, a concern especially in screening scenarios with repeated examinations. Due to recent technological advances, MRI emerges as a potential alternative for lung imaging using 3D steady state free precession and ultra-short echo-time sequences. Therefore, in this study we assessed the performance of three state-of-the-art MRI sequences for the evaluation of pulmonary nodules.

Methods

Lesions of variable sizes were simulated in porcine lungs placed in a dedicated chest phantom mimicking a human thorax, followed by CT and MRI examinations. Two blinded readers evaluated the acquired MR-images locating and measuring every suspect lesion. Using the CT-images as reference, logistic regression was performed to investigate the sensitivity of the tested MRI-sequences for the detection of pulmonary nodules.

Results

For nodules with a diameter of 6 mm, all three sequences achieved high sensitivity values above 0.91. However, the sensitivity dropped for smaller nodules, yielding an average of 0.83 for lesions with 4 mm in diameter and less than 0.69 for lesions with 2 mm in diameter. The positive predictive values ranged between 0.91 and 0.96, indicating a low amount of false positive findings. Furthermore, the size measurements done on the MR-images were subject to a bias ranging from 0.83 mm to -1.77 mm with standard deviations ranging from 1.40 mm to 2.11 mm. There was no statistically significant difference between the three tested sequences.

Conclusion

While showing promising sensitivity values for lesions larger than 4 mm, MRI appears to be not yet suited for lung cancer screening. Nonetheless, the three tested MRI sequences yielded high positive predictive values and accurate size measurements; therefore, MRI could potentially figure as imaging method of the chest in selected follow-up scenarios, e.g. of incidental findings subject to the Fleischner Criteria.

SUBMITTER: Bonert M 

PROVIDER: S-EPMC7757901 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic accuracy of magnetic resonance imaging for the detection of pulmonary nodules simulated in a dedicated porcine chest phantom.

Bonert Madeleine M   Schneider Moritz M   Solyanik Olga O   Hellbach Katharina K   Bondesson David D   Gaass Thomas T   Thaens Natalie N   Ricke Jens J   Benkert Thomas T   Dinkel Julien J  

PloS one 20201223 12


<h4>Objective</h4>CT serves as gold standard for the evaluation of pulmonary nodules. However, CT exposes patients to ionizing radiation, a concern especially in screening scenarios with repeated examinations. Due to recent technological advances, MRI emerges as a potential alternative for lung imaging using 3D steady state free precession and ultra-short echo-time sequences. Therefore, in this study we assessed the performance of three state-of-the-art MRI sequences for the evaluation of pulmon  ...[more]

Similar Datasets

| S-EPMC3744172 | biostudies-literature
| S-EPMC8380068 | biostudies-literature
| S-EPMC6331282 | biostudies-literature
| S-EPMC8252537 | biostudies-literature
| S-EPMC6317805 | biostudies-literature
| S-EPMC5032623 | biostudies-literature
| S-EPMC7478958 | biostudies-literature
| S-EPMC9127105 | biostudies-literature
| S-EPMC5882099 | biostudies-literature
| S-EPMC10101397 | biostudies-literature