Ontology highlight
ABSTRACT: Purpose
The aim of this study was to investigate the correlations of the gamma passing rates (GPR) with the dose-volumetric parameter changes between the original volumetric modulated arc therapy (VMAT) plans and the actual deliveries of the VMAT plans (DV errors). We compared the correlations of the TrueBeam STx system to those of a C-series linac.Methods
A total of 20 patients with head and neck (H&N) cancer were retrospectively selected for this study. For each patient, two VMAT plans with the TrueBeam STx and Trilogy (C-series linac) systems were generated under similar modulation degrees. Both the global and local GPRs with various gamma criteria (3%/3 mm, 2%/2 mm, 2%/1 mm, 1%/2 mm, and 1%/1 mm) were acquired with the 2D dose distributions measured using the MapCHECK2 detector array. During VMAT deliveries, the linac log files of the multi-leaf collimator positions, gantry angles, and delivered monitor units were acquired. The DV errors were calculated with the 3D dose distributions reconstructed using the log files. Subsequently, Spearman's rank correlation coefficients (rs) and the corresponding p values were calculated between the GPRs and the DV errors.Results
For the Trilogy system, the rs values with p < 0.05 showed weak correlations between the GPRs and the DV errors (rs<0.4) whereas for the TrueBeam STx system, moderate or strong correlations were observed (rs?0.4). The DV errors in the V20Gy of the left parotid gland and those in the mean dose of the right parotid gland showed strong correlations (always with rs > 0.6) with the GPRs with gamma criteria except 3%/3 mm. As the GPRs increased, the DV errors decreased.Conclusion
The GPRs showed strong correlations with some of the DV errors for the VMAT plans for H&N cancer with the TrueBeam STx system.
SUBMITTER: Park JM
PROVIDER: S-EPMC7771856 | biostudies-literature | 2020
REPOSITORIES: biostudies-literature
Park Jong Min JM Choi Chang Heon CH Wu Hong-Gyun HG Kim Jung-In JI
PloS one 20201229 12
<h4>Purpose</h4>The aim of this study was to investigate the correlations of the gamma passing rates (GPR) with the dose-volumetric parameter changes between the original volumetric modulated arc therapy (VMAT) plans and the actual deliveries of the VMAT plans (DV errors). We compared the correlations of the TrueBeam STx system to those of a C-series linac.<h4>Methods</h4>A total of 20 patients with head and neck (H&N) cancer were retrospectively selected for this study. For each patient, two VM ...[more]