Unknown

Dataset Information

0

A pilot with a helicopter in the left atrium: a case report of an embolized Watchman device


ABSTRACT: Abstract

Background

Embolization of a Watchman device in patients undergoing percutaneous left atrial appendage (LAA) closure is a rare complication. Retrieval of the device can be achieved percutaneously with a snaring technique or a bioptome.

Case summary

We present an illustrative case of Watchman closure device embolization during an LAA closure attempt in a 77-year-old man. The complication was caused by anatomical restraints of the LAA, in particular limited depth resulting in poor stability of the closure device after implantation. By the use of a double snaring technique, however, we managed to retrieve the device percutaneously. By applying such strategy, a more invasive surgical approach can be prevented.

Discussion

Prompt response is needed when embolization of a Watchman device occurs. A snaring technique can be used in most cases to attempt device retrieval. A double snaring technique is in this case associated with higher retrieval success than a single snaring technique.

SUBMITTER: Halim J 

PROVIDER: S-EPMC7780468 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC6693090 | biostudies-literature
| S-EPMC5708950 | biostudies-literature
| S-EPMC8759468 | biostudies-literature
| S-EPMC7501917 | biostudies-literature
| S-EPMC7780441 | biostudies-literature
| S-EPMC8114721 | biostudies-literature
| S-EPMC8645954 | biostudies-literature
| S-EPMC7411216 | biostudies-literature
| S-EPMC6764553 | biostudies-other
| S-EPMC6739757 | biostudies-literature