The gold standard for COVID-19 diagnosis is currently a real-time reverse transcriptase polymerase chain reaction (RT-PCR) to detect SARS-CoV-2. This is most commonly performed on respiratory secretions obtained via a nasopharyngeal swab. Due to supply chain limitations and high demand worldwide because of the COVID-19 pandemic, access to commercial nasopharyngeal swabs has not been assured. 3D printing methods have been used to meet the shortfall. For longer-term considerations, 3D printing may not compare well with injection molding as a production method due to the challenging scalability and greater production costs of 3D printing.Methods
To secure sufficient nasopharyngeal swab availability for our national healthcare system, we designed a novel injection molded nasopharyngeal swab (the IM2 swab). We performed a clinical diagnostic study comparing the IM2 swab to the Copan FLOQSwab. Forty patients with a known diagnosis of COVID-19 and 10 healthy controls were recruited. Paired nasopharyngeal swabs were obtained from the same nostril of each participant and tested for SARS-CoV-2 by RT-PCR.Results
When compared to the Copan FLOQswab, results from the IM2 swab displayed excellent overall agreement and positive percent agreement of 96.0% and 94.9%, respectively. There was no significant difference in mean RT-PCR cycle threshold values for the ORF1ab (28.05 vs. 28.03, p = 0.97) and E-gene (29.72 vs. 29.37, p = 0.64) targets, respectively. We did not observe any significant adverse events and there was no significant difference in patient-reported pain.Conclusion
In summary, the IM2 nasopharyngeal swab is a clinically safe, highly accurate option to commercial nasopharyngeal swabs.
SUBMITTER: Tay JK
PROVIDER: S-EPMC7799401 | biostudies-literature |
REPOSITORIES: biostudies-literature