ABSTRACT: Background and purpose
An audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy was developed and tested through a multinational research group and performed on a national level in five different countries.Materials and methods
Centres participating in this study were asked to provide OFs calculated by their TPSs for 10?×?10?cm2, 6?×?6?cm2, 4?×?4?cm2, 3?×?3?cm2 and 2?×?2?cm2 field sizes using an SSD of 100?cm. The ratio of these calculated OFs to reference OFs was analysed. The action limit was ±3% for the 2?×?2?cm2 field and ±2% for all other fields.Results
OFs for more than 200 different beams were collected in total. On average, the OFs for small fields calculated by TPSs were generally larger than measured reference data. These deviations increased with decreasing field size. On a national level, 30% and 31% of the calculated OFs of the 2?×?2?cm2 field exceeded the action limit of 3% for nominal beam energies of 6?MV and for nominal beam energies higher than 6 MV, respectively.Conclusion
Modern TPS beam models generally overestimate the OFs for small fields. The verification of calculated small field OFs is a vital step and should be included when commissioning a TPS. The methodology outlined in this study can be used to identify potential discrepancies in clinical beam models.
SUBMITTER: Lechner W
PROVIDER: S-EPMC7807586 | biostudies-literature | 2018 Jan
REPOSITORIES: biostudies-literature
Lechner Wolfgang W Wesolowska Paulina P Azangwe Godfrey G Arib Mehenna M Alves Victor Gabriel Leandro VGL Suming Luo L Ekendahl Daniela D Bulski Wojciech W Samper José Luis Alonso JLA Vinatha Sumanth Panyam SP Siri Srimanoroth S Tomsej Milan M Tenhunen Mikko M Povall Julie J Kry Stephen F SF Followill David S DS Thwaites David I DI Georg Dietmar D Izewska Joanna J
Physics and imaging in radiation oncology 20180101
<h4>Background and purpose</h4>An audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy was developed and tested through a multinational research group and performed on a national level in five different countries.<h4>Materials and methods</h4>Centres participating in this study were asked to provide OFs calculated by their TPSs for 10 × 10 cm<sup>2</sup>, 6 × 6 cm<sup>2</sup>, 4 × 4 cm<sup>2</sup>, 3 ...[more]