Unknown

Dataset Information

0

Evaluation of an a priori scatter correction algorithm for cone-beam computed tomography based range and dose calculations in proton therapy.


ABSTRACT:

Background and purpose

Scatter correction of cone-beam computed tomography (CBCT) projections may enable accurate online dose-delivery estimations in photon and proton-based radiotherapy. This study aimed to evaluate the impact of scatter correction in CBCT-based proton range/dose calculations, in scans acquired in both proton and photon gantries.

Material and methods

CBCT projections of a Catphan and an Alderson phantom were acquired on both a proton and a photon gantry. The scatter corrected CBCTs (corrCBCTs) and the clinical reconstructions (stdCBCTs) were compared against CTs rigidly registered to the CBCTs (rigidCTs). The CBCTs of the Catphan phantom were segmented by materials for CT number analysis. Water equivalent path length (WEPL) maps were calculated through the Alderson phantom while proton plans optimized on the rigidCT and recalculated on all CBCTs were compared in a gamma analysis.

Results

In medium and high-density materials, the corrCBCT CT numbers were much closer to those of the rigidCT than the stdCBCTs. E.g. in the 50% bone segmentations the differences were reduced from above 300 HU (with stdCBCT) to around 60-70 HU (with corrCBCT). Differences in WEPL from the rigidCT were typically well below 5 mm for the corrCBCTs, compared to well above 10 mm for the stdCBCTs with the largest deviations in the head and thorax regions. Gamma pass rates (2%/2mm) when comparing CBCT-based dose re-calculations to rigidCT calculations were improved from around 80% (with stdCBCT) to mostly above 90% (with corrCBCT).

Conclusion

Scatter correction leads to substantial artefact reductions, improving accuracy of CBCT-based proton range/dose calculations.

SUBMITTER: Andersen AG 

PROVIDER: S-EPMC7807858 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of an <i>a priori</i> scatter correction algorithm for cone-beam computed tomography based range and dose calculations in proton therapy.

Andersen Andreas Gravgaard AG   Park Yang-Kyun YK   Elstrøm Ulrik Vindelev UV   Petersen Jørgen Breede Baltzer JBB   Sharp Gregory C GC   Winey Brian B   Dong Lei L   Muren Ludvig Paul LP  

Physics and imaging in radiation oncology 20201027


<h4>Background and purpose</h4>Scatter correction of cone-beam computed tomography (CBCT) projections may enable accurate online dose-delivery estimations in photon and proton-based radiotherapy. This study aimed to evaluate the impact of scatter correction in CBCT-based proton range/dose calculations, in scans acquired in both proton and photon gantries.<h4>Material and methods</h4>CBCT projections of a Catphan and an Alderson phantom were acquired on both a proton and a photon gantry. The scat  ...[more]

Similar Datasets

| S-EPMC6684491 | biostudies-literature
| S-EPMC5619659 | biostudies-literature
| S-EPMC7807872 | biostudies-literature
| S-EPMC4508229 | biostudies-literature
| S-EPMC8321184 | biostudies-literature
| S-EPMC6131388 | biostudies-literature
| S-EPMC9043180 | biostudies-literature
| S-EPMC10023538 | biostudies-literature
| S-EPMC5849848 | biostudies-other
| S-EPMC7742212 | biostudies-literature