Comparison of paediatric King Vision™ videolaryngoscope and Macintosh laryngoscope for elective tracheal intubation in children of age less than 1 year: A randomised clinical trial.


ABSTRACT:

Background and aims

Paediatric airway, because of its consistent anatomical differences from that of an adult, often encounters difficulty in aligning the line of sight with the laryngeal inlet during intubation. Paediatric videolaryngoscopes (VLs), by obviating the need for aligning the line of sight with the glottis, offer several advantages over direct laryngoscopy. Therefore, this study aimed to compare the recently introduced paediatric King Vision™ VL (KVL) and the direct laryngoscope with Macintosh blade for elective tracheal intubation in infants of age <1 year.

Methods

Seventy-eight infants of American Society of Anesthesiologists physical status 1 and 2, scheduled for elective surgery, were enrolled for this prospective randomised clinical trial and randomised into either of the two groups - Group K and Group C, where the infants were intubated using size 1 King Vision or direct laryngoscope with Macintosh blade. The primary objective of this study was the time taken for intubation and the first-attempt intubation success rate.

Results

Time to intubate (25.90 ± 2.34 s vs. 25.03 ± 1.42 s, P = 0.05) and first-attempt intubation success rate (100% vs. 100%, P = 1) were similar between the groups, whereas glottic visualisation (P = 0.01), alternate techniques used to assist intubation (P < 0.001), the ease of intubation (P = 0.02) and intubation difficulty score (P = 0.01) were better in Group K than that in Group C.

Conclusion

The outcome of KVL and Macintosh laryngoscope was similar in terms of time taken for intubation and first-attempt intubation success rate with KVL having superior glottic visualisation, better ease of intubation and lower intubation difficulty score for elective intubations in children of age <1 year.

SUBMITTER: Manirajan M 

PROVIDER: S-EPMC7815006 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of paediatric King Vision™ videolaryngoscope and Macintosh laryngoscope for elective tracheal intubation in children of age less than 1 year: A randomised clinical trial.

Manirajan Manov M   Bidkar Prasanna Udupi PU   Sivakumar Ranjith Kumar RK   Lata Suman S   Srinivasan Gnanasekaran G   Jha Ajay Kumar AK  

Indian journal of anaesthesia 20201101 11


<h4>Background and aims</h4>Paediatric airway, because of its consistent anatomical differences from that of an adult, often encounters difficulty in aligning the line of sight with the laryngeal inlet during intubation. Paediatric videolaryngoscopes (VLs), by obviating the need for aligning the line of sight with the glottis, offer several advantages over direct laryngoscopy. Therefore, this study aimed to compare the recently introduced paediatric King Vision™ VL (KVL) and the direct laryngosc  ...[more]

Similar Datasets

| S-EPMC9391737 | biostudies-literature
| S-EPMC6717380 | biostudies-literature
| S-EPMC7429469 | biostudies-literature
| S-EPMC8234758 | biostudies-literature
| S-EPMC5578637 | biostudies-literature
| S-EPMC4575358 | biostudies-literature
| S-EPMC5416723 | biostudies-literature
| S-EPMC10863731 | biostudies-literature
| S-EPMC7392330 | biostudies-literature
| S-EPMC4110170 | biostudies-literature