Evaluation of daily online contour adaptation by radiation therapists for prostate cancer treatment on an MRI-guided linear accelerator.


ABSTRACT:

Background and purpose

Magnetic resonance (MR)-guided linear accelerator (MR-Linac) systems have changed radiotherapy workflows. The addition of daily online contour adaptation allows for higher precision treatment, but also increases the workload of those involved. We train radiation therapists (RTTs) to perform daily online contour adaptation for MR-Linac treatment of prostate cancer (PCa) patients. The purpose of this study was to evaluate these prostate contours by performing an interfraction and interobserver analysis.

Materials and methods

Clinical target volume (CTV) contours generated online by RTTs from 30 low-intermediate risk PCa patients, treated with 5x7.25 Gy, were used. Two physicians (Observers) judged the RTTs contours and performed adaptations when necessary. Interfraction relative volume differences between the first and the subsequent fractions were calculated for the RTTs, Observer 1, and Observer 2. Additionally, interobserver dice's similarity coefficient (DSC) for fraction 2-5 was calculated with the RTTs- and physician-adapted contours. Clinical acceptability of the RTTs contours was judged by a third observer.

Results

Mean (SD) online contour adaptation time was 12.6 (±3.8) minutes and overall median (interquartile range [IQR]) relative volume difference was 9.3% (4.4-13.0). Adaptations by the observers were mostly performed at the apex and base of the prostate. Median (IQR) interobserver DSC between RTTs and Observer 1, RTTs and Observer 2, and Observer 1 and 2 was 0.99 (0.98-1.00), 1.00 (0.98-1.00), and 1.00 (0.99-1.00), respectively. Contours were acceptable for clinical use in 113 (94.2%) fractions. Dose-volume histogram (DVH) analysis showed significant CTV underdosage for one of the seven identified outliers.

Conclusion

Daily online contour adaptation by RTTs is clinically feasible for MR-Linac treatment of PCa.

SUBMITTER: Willigenburg T 

PROVIDER: S-EPMC7822780 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of daily online contour adaptation by radiation therapists for prostate cancer treatment on an MRI-guided linear accelerator.

Willigenburg Thomas T   de Muinck Keizer Daan M DM   Peters Max M   Claes An A   Lagendijk Jan J W JJW   de Boer Hans C J HCJ   van der Voort van Zyp Jochem R N JRN  

Clinical and translational radiation oncology 20210114


<h4>Background and purpose</h4>Magnetic resonance (MR)-guided linear accelerator (MR-Linac) systems have changed radiotherapy workflows. The addition of daily online contour adaptation allows for higher precision treatment, but also increases the workload of those involved. We train radiation therapists (RTTs) to perform daily online contour adaptation for MR-Linac treatment of prostate cancer (PCa) patients. The purpose of this study was to evaluate these prostate contours by performing an inte  ...[more]

Similar Datasets

| S-EPMC9460551 | biostudies-literature
| S-EPMC9838213 | biostudies-literature
| S-EPMC8057955 | biostudies-literature
| S-EPMC9280040 | biostudies-literature
| S-EPMC9280019 | biostudies-literature
| S-EPMC10104451 | biostudies-literature
| S-EPMC8867050 | biostudies-literature
| S-EPMC8985054 | biostudies-literature
| S-EPMC5711244 | biostudies-literature