The concept of a cementless isoelastic monoblock cup made of highly cross-linked polyethylene infused with vitamin E: radiological analyses of migration and wear using EBRA and clinical outcomes at mid-term follow-up.


ABSTRACT:

Background

The newest generation of cementless titanium-coated, isoelastic monoblock cup with vitamin E-blended highly cross-linked polyethylene (HXLPE) was introduced to the market in 2009. The aim of the present study was to obtain mid-term follow-up data including migration and wear analyses.

Methods

This prospective study investigated 101 primary total hip arthroplasty (THA) cases in 96 patients treated at a single institution. Patients were allowed full weight-bearing on the first day postoperatively. Harris hip score (HHS) and pain and satisfication on a visual analogue scale (VAS) were assessed at a mean follow-up of 79.0?months. Migration and wear were assessed using Einzel-Bild-Roentgen-Analyse (EBRA) software. Radiological acetabular bone alterations and complications were documented.

Results

At mid-term follow-up (mean 79.0?months, range: 51.8-101.7), 81 cases with complete clinical and radiological data were analyzed. Utilisable EBRA measurements were obtained for 42 hips. The mean HHS was 91.1 (range 38.0-100.0), VAS satisfaction was 9.6 (range 6.0-10.0), VAS rest pain was 0.2 (range 0.0-4.0), and VAS load pain was 0.6 (range 0.0-9.0). Mean migration was 0.86?mm (range: 0.0-2.56) at 24?months and 1.34?mm (range: 0.09-3.14) at 5?years, and the mean annual migration rate was 0.22 (range: -?0.24-1.34). The mean total wear was 0.4?mm (range: 0.03-1.0), corresponding to a mean annual wear rate of 0.06?mm per year (range: 0.0-0.17). Radiographic analysis did not reveal any cases of osteolysis, and no revision surgeries had to be performed.

Conclusions

After using vitamin-E blended HXLPE in cementless isoelastic monoblock cups, there were no obvious signs of osteolysis or aseptic loosening occurred. No patients required revision surgery after mid-term follow-up. Cup migration and wear values were well below the benchmarks considered predictive for potential future failure.

Trial registration

The trial registration number on ClinicalTrials.gov : NCT04322916 (retrospectively registered at 26.03.2020).

SUBMITTER: Afghanyar Y 

PROVIDER: S-EPMC7827971 | biostudies-literature | 2021 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

The concept of a cementless isoelastic monoblock cup made of highly cross-linked polyethylene infused with vitamin E: radiological analyses of migration and wear using EBRA and clinical outcomes at mid-term follow-up.

Afghanyar Yama Y   Joser Sebastian S   Tecle Jonas J   Drees Philipp P   Dargel Jens J   Rehbein Philipp P   Kutzner Karl Philipp KP  

BMC musculoskeletal disorders 20210123 1


<h4>Background</h4>The newest generation of cementless titanium-coated, isoelastic monoblock cup with vitamin E-blended highly cross-linked polyethylene (HXLPE) was introduced to the market in 2009. The aim of the present study was to obtain mid-term follow-up data including migration and wear analyses.<h4>Methods</h4>This prospective study investigated 101 primary total hip arthroplasty (THA) cases in 96 patients treated at a single institution. Patients were allowed full weight-bearing on the  ...[more]

Similar Datasets

| S-EPMC2269018 | biostudies-other
| S-EPMC8158183 | biostudies-literature
| S-EPMC8013815 | biostudies-literature
| S-EPMC5119444 | biostudies-literature
| S-EPMC8023900 | biostudies-literature
| S-EPMC4564778 | biostudies-literature
| S-EPMC6366473 | biostudies-literature
| S-EPMC8199669 | biostudies-literature
| S-EPMC9726741 | biostudies-literature
| S-EPMC3461862 | biostudies-literature