Unknown

Dataset Information

0

Comparison of neurofilament light chain results between two independent facilities.


ABSTRACT:

Objectives

To examine levels of neurofilament light chain (NFL) in identical cerebrospinal fluid (CSF) and blood samples at two different facilities, and how differences affect interpretation of levels within and above the normal range.

Methods

CSF and plasma from patients with multiple sclerosis (MS) and healthy controls (HCs) were analysed by Simoa (Quanterix) for levels of NFL providing a total of 165 CSF samples (119 from MS) and 225 plasma samples (180 from MS).

Results

CSF and plasma concentrations highly correlated between NFL laboratory facilities (R: 0.92 and 0.84, <0.0001, respectively), and there were no differences between facilities. A bias between the two sites for plasma was -0.95?pg/mL and for CSF -73.53?pg/mL. The cut-offs for CSF were 807.5 and 571.0?pg/mL at site 1 and site 2, respectively; the cut-offs for plasma were 13.0 and 11.8?pg/mL, respectively. Seven out of 180 plasma samples (3.9%) and 3 out of 119 CSF samples (2.5%) from MS patients could be reclassified as normal/abnormal, that is, below/above cut-off, when measured at different facilities.

Conclusion

Our study demonstrates that results of NFL in CSF and blood measured with SIMOA are comparable between facilities. Nevertheless, healthcare practitioners should consider reference values at different laboratories, since different sensitivity/specificity can affect interpretation when low values are adjacent to cut-offs.

SUBMITTER: Sejbaek T 

PROVIDER: S-EPMC7871722 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of neurofilament light chain results between two independent facilities.

Sejbaek Tobias T   Mendoza Jason P JP   Penner Natasha N   Madsen Jonna Skov JS   Olsen Dorte Aalund DA   Illes Zsolt Z  

BMJ neurology open 20201019 2


<h4>Objectives</h4>To examine levels of neurofilament light chain (NFL) in identical cerebrospinal fluid (CSF) and blood samples at two different facilities, and how differences affect interpretation of levels within and above the normal range.<h4>Methods</h4>CSF and plasma from patients with multiple sclerosis (MS) and healthy controls (HCs) were analysed by Simoa (Quanterix) for levels of NFL providing a total of 165 CSF samples (119 from MS) and 225 plasma samples (180 from MS).<h4>Results</h  ...[more]

Similar Datasets

| S-EPMC8761779 | biostudies-literature
| S-EPMC4999594 | biostudies-literature
| S-EPMC8758474 | biostudies-literature
| S-EPMC7672085 | biostudies-literature
| S-EPMC4456658 | biostudies-literature
| S-EPMC6043776 | biostudies-literature
| S-EPMC4531057 | biostudies-literature
| S-EPMC7500015 | biostudies-literature
| S-EPMC9266981 | biostudies-literature
| S-EPMC6954884 | biostudies-literature