Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection.


ABSTRACT:

Background

Two automatable in-house protocols for high-troughput RNA extraction from nasopharyngeal swabs for SARS-CoV-2 detection have been evaluated.

Methods

One hundred forty one SARS-CoV-2 positive samples were collected during a period of 10-days. In-house protocols were based on extraction with magnetic beads and designed to be used with either the Opentrons OT-2 (OT-2in-house) liquid handling robot or the MagMAXTM Express-96 system (MMin-house). Both protocols were tested in parallel with a commercial kit that uses the MagMAXTM system (MMkit). Nucleic acid extraction efficiencies were calculated from a SARS-CoV-2 DNA positive control.

Results

No significant differences were found between both in-house protocols and the commercial kit in their performance to detect positive samples. The MMkit was the most efficient although the MMin-house presented, in average, lower Cts than the other two. In-house protocols allowed to save between 350€ and 400€ for every 96 extracted samples compared to the commercial kit.

Conclusion

The protocols described harness the use of easily available reagents and an open-source liquid handling system and are suitable for SARS-CoV-2 detection in high throughput facilities.

SUBMITTER: Lazaro-Perona F 

PROVIDER: S-EPMC7886139 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of two automated low-cost RNA extraction protocols for SARS-CoV-2 detection.

Lázaro-Perona Fernando F   Rodriguez-Antolín Carlos C   Alguacil-Guillén Marina M   Gutiérrez-Arroyo Almudena A   Mingorance Jesús J   García-Rodriguez Julio J  

PloS one 20210216 2


<h4>Background</h4>Two automatable in-house protocols for high-troughput RNA extraction from nasopharyngeal swabs for SARS-CoV-2 detection have been evaluated.<h4>Methods</h4>One hundred forty one SARS-CoV-2 positive samples were collected during a period of 10-days. In-house protocols were based on extraction with magnetic beads and designed to be used with either the Opentrons OT-2 (OT-2in-house) liquid handling robot or the MagMAXTM Express-96 system (MMin-house). Both protocols were tested i  ...[more]

Similar Datasets

| S-EPMC8777774 | biostudies-literature
| S-EPMC7599812 | biostudies-literature
| S-EPMC8231976 | biostudies-literature
| S-EPMC8284070 | biostudies-literature
| S-EPMC8529346 | biostudies-literature
| S-EPMC10663557 | biostudies-literature
| S-EPMC8003619 | biostudies-literature
2015-01-07 | GSE64673 | GEO
| S-EPMC8280604 | biostudies-literature
| S-EPMC8568942 | biostudies-literature