ABSTRACT: Background
The performance of the Roche Elecsys® Anti-SARS-CoV-2, Abbott Architect SARS-CoV-2 IgM, Abbott Architect SARS-CoV-2 IgG, Euroimmun SARS-CoV-2 IgA, Euroimmun SARS-CoV-2 IgG ELISA, and Trillium IgG/IgM rapid assays was evaluated in Jamaica.Methods
Diagnostic sensitivities of the assays were assessed by testing serum samples from SARS-CoV-2 PCR-confirmed persons and diagnostic specificity was assessed by testing serum samples collected during 2018-2019 from healthy persons and from persons with antibodies to a wide range of viral infections.Results
Serum samples collected ≥14 days after onset of symptoms, or an initial SARS-CoV-2 RT-PCR positive test for asymptomatics, showed diagnostic sensitivities ranging from 67.9 to 75.0% when including all possible disease severities and increased to 90.0-95.0% when examining those with moderate to critical disease. Grouping moderate to critical disease showed a significant association with a SARS-CoV-2 antibody positive result for all assays. Diagnostic specificity ranged from 96.7 to 100.0%. For all assays examined, SARS-CoV-2 real-time PCR cycle threshold (Ct) values of the initial nasopharyngeal swab sample testing positive were significantly different for samples testing antibody positive versus negative.Conclusions
These data from a predominantly African descent Caribbean population show comparable diagnostic sensitivities and specificities for all testing platforms assessed and limited utility of these tests for persons with asymptomatic and mild infections.