ABSTRACT: Objective
To evaluate the accuracy of respiratory mechanics using dynamic signal analysis during noninvasive pressure support ventilation (PSV).Methods
A Respironics V60 ventilator was connected to an active lung simulator to model normal, restrictive, obstructive, and mixed obstructive and restrictive profiles. The PSV was adjusted to maintain tidal volumes (VT) that achieved 5.0, 7.0, and 10.0?mL/kg body weight, and the positive end-expiration pressure (PEEP) was set to 5?cmH2O. Ventilator performance was evaluated by measuring the flow, airway pressure, and volume. The system compliance (Crs) and airway resistance (inspiratory and expiratory resistance, Rinsp and Rexp, respectively) were calculated.Results
Under active breathing conditions, the Crs was overestimated in the normal and restrictive models, and it decreased with an increasing pressure support (PS) level. The Rinsp calculated error was approximately 10% at 10.0?mL/kg of VT, and similar results were obtained for the calculated Rexp at 7.0?mL/kg of VT.Conclusion
Using dynamic signal analysis, appropriate tidal volume was beneficial for Rrs, especially for estimating Rexp during assisted ventilation. The Crs measurement was also relatively accurate in obstructive conditions.
SUBMITTER: Chen Y
PROVIDER: S-EPMC7925949 | biostudies-literature | 2021 Feb
REPOSITORIES: biostudies-literature
Chen Yuqing Y Yuan Yueyang Y Zhang Hai H Li Feng F Zhou Xin X
The Journal of international medical research 20210201 2
<h4>Objective</h4>To evaluate the accuracy of respiratory mechanics using dynamic signal analysis during noninvasive pressure support ventilation (PSV).<h4>Methods</h4>A Respironics V60 ventilator was connected to an active lung simulator to model normal, restrictive, obstructive, and mixed obstructive and restrictive profiles. The PSV was adjusted to maintain tidal volumes (V<sub>T</sub>) that achieved 5.0, 7.0, and 10.0 mL/kg body weight, and the positive end-expiration pressure (PEEP) was set ...[more]