ABSTRACT: Background and purpose
Despite the improvement in technology, endovascular treatment of bifurcation intracranial wide-neck aneurysms remains challenging, mainly due to the difficulty of maintaining coils within the aneurysm sac without compromising the patency of bifurcation arteries. The Woven EndoBridge (WEB) device is a recent intrasaccular braided device specifically dedicated to treating such aneurysms with a wide neck by disrupting the flow in the aneurysmal neck and promoting progressive aneurysmal thrombosis.Materials and methods
Using several health data bases, we conducted a systematic review of all published studies of WEB endovascular treatment in intracranial aneurysms from 2010 onward to evaluate its efficacy and safety profile.Results
The literature search identified 6 relevant studies (7 articles) including wide-neck bifurcation aneurysms in ≥80% of cases. Clinical data supporting the efficacy and safety of the WEB are limited to noncomparative cohort studies with large heterogeneity from a methodologic standpoint. The WEB deployment was feasible with a success rate of 93%-100%. Permanent morbidity (mRS of >1 at last follow-up) and mortality were measured at 2.2%-6.7% and 0%-17%, respectively. The adequate occlusion rate (total occlusion or neck remnant) varied between 65% and 85.4% at midterm follow-up (range, 3.3-27.4 months).Conclusions
Endovascular treatment of bifurcation wide-neck aneurysms with the WEB device is feasible and allows an acceptably adequate aneurysm occlusion rate; however, the rate of neck remnants is not negligible. The WEB device needs further clinical and anatomic evaluation with long-term prospective studies, especially of the risk of WEB compression. Prospective controlled studies should be encouraged.
SUBMITTER: Armoiry X
PROVIDER: S-EPMC7960317 | biostudies-literature | 2016 May
REPOSITORIES: biostudies-literature

Armoiry X X Turjman F F Hartmann D J DJ Sivan-Hoffmann R R Riva R R Labeyrie P E PE Aulagner G G Gory B B
AJNR. American journal of neuroradiology 20151119 5
<h4>Background and purpose</h4>Despite the improvement in technology, endovascular treatment of bifurcation intracranial wide-neck aneurysms remains challenging, mainly due to the difficulty of maintaining coils within the aneurysm sac without compromising the patency of bifurcation arteries. The Woven EndoBridge (WEB) device is a recent intrasaccular braided device specifically dedicated to treating such aneurysms with a wide neck by disrupting the flow in the aneurysmal neck and promoting prog ...[more]