ABSTRACT: Background and purpose
The mismatch between lesions identified in perfusion- and diffusion-weighted MR imaging is typically used to identify tissue at risk of infarction in acute stroke. The purpose of this study was to analyze the variability of mismatch volumes resulting from different time-to-peak or time-to-maximum estimation techniques used for hypoperfused tissue definition.Materials and methods
Data of 50 patients with middle cerebral artery stroke and intracranial vessel occlusion imaged within 6 hours of symptom onset were analyzed. Therefore, 10 different TTP/Tmax techniques and delay thresholds between +2 and +12 seconds were used for calculation of perfusion lesions. Diffusion lesions were semiautomatically segmented and used for mismatch quantification after registration.Results
Mean volumetric differences up to 40 and 100 mL in individual patients were found between the mismatch volumes calculated by the 10 TTP/Tmax estimation techniques for typically used delay thresholds. The application of typical criteria for the identification of patients with a clinically relevant mismatch volume resulted in different mismatch classifications in ≤24% of all cases, depending on the TTP/Tmax estimation method used.Conclusions
High variations of tissue-at-risk volumes have to be expected when using different TTP/Tmax estimation techniques. An adaption of different techniques by using correction formulas may enable more comparable study results until a standard has been established by agreement.