Ontology highlight
ABSTRACT: Background
The authors developed a cross-laser projection system (CLP) to place a femoral neck-sparing short stem using the minimally invasive anterolateral supine approach in total hip arthroplasty. This study aimed to verify the utility of CLP.Methods
Thirty joints were assessed with the MiniHip (Corin). The authors compared femoral component implantation with a patient-specific femoral osteotomy guide (PSG) for the femoral neck-cut (PSG group), with the CLP attached to the rasp handle to irradiate the cross-laser to the target of PSG (CLP group), and without PSG or CLP (control group).Results
In the CLP group, the positional deviation of anteversion, anterior/posterior tilt and varus/valgus placement of the stem postoperatively were 1.8° ± 0.2°, 2.0° ± 2.0° and 2.0° ± 0.1°, respectively. The positional deviation of anteversion (p < 0.001) and anterior/posterior tilt (p = 0.036) were significantly smaller than those in the other groups.Conclusions
CLP improves the accuracy of MiniHip femoral prosthesis placement.
SUBMITTER: Fujii H
PROVIDER: S-EPMC7988605 | biostudies-literature |
REPOSITORIES: biostudies-literature