ABSTRACT: Background and purpose
Few studies on magnetic resonance imaging (MRI) only head and neck radiation treatment planning exist, and none using a generally available software. The aim of this study was to evaluate the accuracy of absorbed dose for head and neck synthetic computed tomography data (sCT) generated by a commercial convolutional neural network-based algorithm.Materials and methods
For 44 head and neck cancer patients, sCT were generated and the geometry was validated against computed tomography data (CT). The clinical CT based treatment plan was transferred to the sCT and recalculated without re-optimization, and differences in relative absorbed dose were determined for dose-volume-histogram (DVH) parameters and the 3D volume.Results
For overall body, the results of the geometric validation were (Mean ± 1sd): Mean error -5 ± 10 HU, mean absolute error 67 ± 14 HU, Dice similarity coefficient 0.98 ± 0.05, and Hausdorff distance difference 4.2 ± 1.7 mm. Water equivalent depth difference for region Th1-C7, mid mandible and mid nose were -0.3 ± 3.4, 1.1 ± 2.0 and 0.7 ± 3.8 mm respectively. The maximum mean deviation in absorbed dose for all DVH parameters was 0.30% (0.12 Gy). The absorbed doses were considered equivalent (p-value < 0.001) and the mean 3D gamma passing rate was 99.4 (range: 95.7-99.9%).Conclusions
The convolutional neural network-based algorithm generates sCT which allows for accurate absorbed dose calculations for MRI-only head and neck radiation treatment planning. The sCT allows for statistically equivalent absorbed dose calculations compared to CT based radiotherapy.