Differences between 3D isovoxel fat suppression VIBE MRI and CT models of proximal femur osseous anatomy: A preliminary study for bone tumor resection planning.


ABSTRACT:

Purpose

To evaluate the osseous anatomy of the proximal femur extracted from a 3D-MRI volumetric interpolated breath-hold (VIBE) sequence using either a Dixon or water excitation (WE) fat suppression method, and to measure the overall difference using CT as a reference standard.

Material and methods

This retrospective study reviewed imaging of adult patients with hip pain who underwent 3D hip MRI and CT. A semi-automatically segmented CT model served as the reference standard, and MRI segmentation was performed manually for each unilateral hip joint. The differences between Dixon-VIBE-3D-MRI vs. CT, and WE-VIBE-3D-MRI vs. CT, were measured. Equivalence tests between Dixon-VIBE and WE-VIBE models were performed with a threshold of 0.1 mm. Bland-Altman plots and Lin's concordance-correlation coefficient were used to analyze the agreement between WE and Dixon sequences. Subgroup analyses were performed for the femoral head/neck, intertrochanteric, and femoral shaft areas.

Results

The mean and maximum differences between Dixon-VIBE-3D-MRI vs. CT were 0.2917 and 3.4908 mm, respectively, whereas for WE-VIBE-3D-MRI vs. CT they were 0.3162 and 3.1599 mm. The mean differences of the WE and Dixon methods were equivalent (P = 0.0292). However, the maximum difference was not equivalent between the two methods and it was higher in WE method. Lin's concordance-correlation coefficient showed poor agreement between Dixon and WE methods. The mean differences between the CT and 3D-MRI models were significantly higher in the femoral shaft area (P = 0.0004 for WE and P = 0.0015 for Dixon) than in the other areas. The maximum difference was greatest in the intertrochanteric area for both techniques.

Conclusion

The difference between 3D-MR and CT models were acceptable with a maximal difference below 3.5mm. WE and Dixon fat suppression methods were equivalent. The mean difference was highest at the femoral shaft area, which was off-center from the magnetization field.

SUBMITTER: Chee CG 

PROVIDER: S-EPMC8087022 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC6588803 | biostudies-other
| S-EPMC9881341 | biostudies-literature
| S-EPMC8225183 | biostudies-literature
| S-EPMC8716646 | biostudies-literature
| S-EPMC9627078 | biostudies-literature
| S-EPMC6910626 | biostudies-literature
| S-EPMC7812921 | biostudies-literature
| S-EPMC8726491 | biostudies-literature
| S-EPMC3270162 | biostudies-literature
| S-EPMC7096175 | biostudies-literature