Development and Evaluation of a Quantitative Fluorescent Lateral Flow Immunoassay for Cystatin-C, a Renal Dysfunction Biomarker.
Ontology highlight
ABSTRACT: The diagnosis, prognosis, and control of chronic kidney disease rely on an understanding of the glomerular filtration rate (GFR). The renal clearance of the cystatin-C is closely associated with the GFR. Cystatin-C is a more suitable GFR marker than the commonly used creatinine. General techniques for cystatin-C calculation, such as particle-enhanced turbidimetric and nephelometric assay, are time-consuming and tedious. Here, we propose a rapid, quantitative immunoassay for the detection of cystatin-C. A fluorescence-based lateral-flow kit was developed in a sandwich format by using a monoclonal antibody. A Linear calibration was obtained over the clinical diagnostic range of 0.023-32 µg/mL and the limit of detection (LOD) was 0.023 µg/mL and the limit of quantification (LOQ) was 0.029 µg/mL. Average recoveries from spiked urine samples ranged from 96-100% and the coefficient of variation was less than 4% for both intra and inter-day assays with excellent repeatability. With the comparison with an ELISA kit, the developed kit is highly sensitive, performs well over the detection range, provides repeatable results in a short time, and can easily be used at point-of-care (POC), making it an ideal candidate for rapid testing in early detection, community screening for renal function disorders.