Unknown

Dataset Information

0

Development and Evaluation of a Quantitative Fluorescent Lateral Flow Immunoassay for Cystatin-C, a Renal Dysfunction Biomarker.


ABSTRACT: The diagnosis, prognosis, and control of chronic kidney disease rely on an understanding of the glomerular filtration rate (GFR). The renal clearance of the cystatin-C is closely associated with the GFR. Cystatin-C is a more suitable GFR marker than the commonly used creatinine. General techniques for cystatin-C calculation, such as particle-enhanced turbidimetric and nephelometric assay, are time-consuming and tedious. Here, we propose a rapid, quantitative immunoassay for the detection of cystatin-C. A fluorescence-based lateral-flow kit was developed in a sandwich format by using a monoclonal antibody. A Linear calibration was obtained over the clinical diagnostic range of 0.023-32 µg/mL and the limit of detection (LOD) was 0.023 µg/mL and the limit of quantification (LOQ) was 0.029 µg/mL. Average recoveries from spiked urine samples ranged from 96-100% and the coefficient of variation was less than 4% for both intra and inter-day assays with excellent repeatability. With the comparison with an ELISA kit, the developed kit is highly sensitive, performs well over the detection range, provides repeatable results in a short time, and can easily be used at point-of-care (POC), making it an ideal candidate for rapid testing in early detection, community screening for renal function disorders.

SUBMITTER: Natarajan S 

PROVIDER: S-EPMC8125764 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7918919 | biostudies-literature
| S-EPMC6189908 | biostudies-literature
| S-EPMC8722402 | biostudies-literature
| S-EPMC6956851 | biostudies-literature
| S-EPMC3000338 | biostudies-literature
| S-EPMC8015349 | biostudies-literature
| S-EPMC8500848 | biostudies-literature
| S-EPMC9174064 | biostudies-literature
| S-EPMC5703415 | biostudies-literature
| S-EPMC7127147 | biostudies-literature