Unknown

Dataset Information

0

Responsive manipulation of neural circuit pathology by fully implantable, front-end multiplexed embedded neuroelectronics.


ABSTRACT: Responsive neurostimulation is increasingly required to probe neural circuit function and treat neuropsychiatric disorders. We introduce a multiplex-then-amplify (MTA) scheme that, in contrast to current approaches (which necessitate an equal number of amplifiers as number of channels), only requires one amplifier per multiplexer, significantly reducing the number of components and the size of electronics in multichannel acquisition systems. It also enables simultaneous stimulation of arbitrary waveforms on multiple independent channels. We validated the function of MTA by developing a fully implantable, responsive embedded system that merges the ability to acquire individual neural action potentials using conformable conducting polymer-based electrodes with real-time onboard processing, low-latency arbitrary waveform stimulation, and local data storage within a miniaturized physical footprint. We verified established responsive neurostimulation protocols and developed a network intervention to suppress pathological coupling between the hippocampus and cortex during interictal epileptiform discharges. The MTA design enables effective, self-contained, chronic neural network manipulation with translational relevance to the treatment of neuropsychiatric disease.

SUBMITTER: Zhao Z 

PROVIDER: S-EPMC8157942 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC3373993 | biostudies-literature
| S-EPMC9258776 | biostudies-literature
| S-EPMC4880021 | biostudies-literature
| S-EPMC9270064 | biostudies-literature
| S-EPMC5640477 | biostudies-literature
| S-EPMC2001238 | biostudies-literature
| S-EPMC7700015 | biostudies-literature
| S-EPMC9411384 | biostudies-literature
| S-EPMC11291159 | biostudies-literature
| S-EPMC4839392 | biostudies-literature