Evaluation of the quality of mammographic breast positioning: a quality improvement study.


ABSTRACT:

Background

Although there are concerns that inadequate breast positioning in mammographic examinations may lead to cancers being missed, few studies have examined the quality of breast positioning, especially in the Canadian context. Our objective was to assess the quality of breast positioning in mammographic examinations in a Quebec-wide representative sample of technologists.

Methods

This quality improvement study was part of a professional inspection launched by the Ordre des technologues en imagerie médicale, en radio-oncologie et en électrophysiologie médicale du Québec among its members. The inspection was conducted between May and July 2017 on a proportionate stratified random sample of all active technologists certified in mammography in Quebec. Each technologist provided images from 15 consecutive mammographic examinations they performed in the previous 6 months. The quality of positioning was then evaluated by senior technologists using a quality assessment tool specifically developed for this inspection. A technologist was deemed to have failed the professional inspection when at least 7 of the 15 mammographic examinations were scored as critical failures. Proportions were calculated accounting for sampling weights and correction for finite population.

Results

Among the 520 technologists certified in mammography in Quebec, 76 technologists (14.6%) were randomly selected for the professional inspection and contributed images from 1127 mammographic examinations. Thirty-eight technologists (weighted percentage 50.3%, 95% confidence interval [CI] 37.6% to 63.0%) failed the professional inspection. Overall, 492 mammographic examinations (43.7%, 95% CI 38.6% to 48.8%) had at least 1 image scored as a critical failure.

Interpretation

Half of the technologists performing mammographic examinations in Quebec who participated in this study failed the inspection, and a substantial proportion of their mammographic examinations demonstrated critical failures in breast positioning. Overall, our findings are concordant with those of previous studies and highlight the need for additional investigations assessing the quality of breast positioning in mammographic examinations in other jurisdictions.

SUBMITTER: Rouette J 

PROVIDER: S-EPMC8191588 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC6624057 | biostudies-literature
| S-EPMC10468499 | biostudies-literature
| S-EPMC9773568 | biostudies-literature
| S-EPMC3528971 | biostudies-other
| S-EPMC8166859 | biostudies-literature
2010-08-13 | GSE18672 | GEO
| S-EPMC6182121 | biostudies-literature
| S-EPMC7172021 | biostudies-literature
| S-EPMC7325699 | biostudies-literature
| S-EPMC5392962 | biostudies-literature