Microbubble-Facilitated Ultrasound Catheter Ablation Causes Microvascular Damage and Fibrosis.


ABSTRACT: High-intensity ultrasound (US) ablation produces deeper myocardial lesions than radiofrequency ablation. The presence of intravascular microbubble (MB) contrast agents enhances pulsed-wave US ablation via cavitation-related histotripsy, potentially facilitating ablation in persistently perfused/conducting myocardium. US ablation catheters were developed and tested in the presence of MBs using ex vivo and in vivo models. High-frame-rate videomicroscopy and US imaging of gel phantom models confirmed MB destruction by inertial cavitation. MB-facilitated US ablation in an ex vivo perfused myocardium model generated shallow (2 mm) lesions and, in an in vivo murine hindlimb model, reduced perfusion by 42% with perivascular hemorrhage and inflammation, but no myonecrosis.