ABSTRACT: Objectives
Decreasing aerosol leaks are of great interest, especially in the recent era of COVID-19. The aim was to investigate intrapulpal heat development, coolant spray patterns, and the preparation efficiency of speed-increasing contra-angle handpieces with the spray air on (mist) or off (water jet) settings during restorative cavity preparations.Methods
Standard-sized cavities were prepared in 80 extracted intact human molar teeth using diamond cylindrical drills with a 1:5 speed-increasing contra-angle handpiece. A custom-made device maintained the standardized lateral drilling force (3 N) and predetermined depth. Temperatures were measured using intrapulpal thermocouple probes. The four experimental groups were as follows: mist cooling mode at 15 mL/min (AIR15), water jet cooling mode at 15 mL/min (JET15), mist cooling mode at 30 mL/min (AIR30), and water jet cooling mode at 30 mL/min (JET30). The coolant spray pattern was captured using macro-photo imaging.Results
The JET15 group had the highest increase in temperature (ΔT = 6.02 °C), while JET30 (ΔT = 2.24 °C; p < 0.001), AIR15 (ΔT = 3.34 °C; p = 0.042), and AIR30 (ΔT = 2.95 °C; p = 0.003) had significantly lower increases in temperature. Fine mist aerosol was formed in the AIR15 and AIR30 preparations but not in the JET15 and JET30 preparations (p < 0.001). The irrigation mode had no influence on the preparation time (p = 0.672).Conclusions
Water jet irrigation using coolant at 30 mL/min appeared to be the optimal mode. Considering the safe intrapulpal temperatures and the absence of fine mist aerosols, this mode can be recommended for restorative cavity preparations.Clinical significance
To increase infection control in dental practices, the water jet irrigation mode of speed-increasing handpieces with coolant flow rates of 30 mL/min should be considered for restorative cavity preparations.