ABSTRACT: Background
Cerebrospinal fluid shunts in the treatment of hydrocephalus, although associated with clinical benefit, have a high failure rate with repeat computed tomography (CT) imaging resulting in a substantial cumulative radiation dose. Therefore, we sought to develop a whole-body ultralow-dose (ULD) CT protocol for the investigation of shunt malfunction and compare it with the reference standard, plain radiographic shunt series (PRSS).Methods
Following ethical approval, using an anthropomorphic phantom and a human cadaveric ventriculoperitoneal shunt model, a whole-body ULD-CT protocol incorporating two iterative reconstruction (IR) algorithms, pure IR and hybrid IR, including 60% filtered back projection and 40% IR was evaluated in 18 adult patients post new shunt implantation or where shunt malfunction was suspected. Effective dose (ED) and image quality were analysed.Results
ULD-CT permitted a 36% radiation dose reduction (median ED 0.16 mSv, range 0.07-0.17, versus 0.25 mSv (0.06-1.69 mSv) for PRSS (p = 0.002). Shunt visualisation in the thoracoabdominal cavities was improved with ULD-CT with pure IR (p = 0.004 and p = 0.031, respectively) and, in contrast to PRSS, permitted visualisation of the entire shunt course (p < 0.001), the distal shunt entry point and location of the shunt tip in all cases. For shunt complications, ULD-CT had a perfect specificity. False positives (3/22, 13.6%) were observed with PRSS.Conclusions
At a significantly reduced radiation dose, whole body ULD-CT with pure IR demonstrated diagnostic superiority over PRSS in the evaluation of cerebrospinal fluid shunt malfunction.
SUBMITTER: Ryan DJ
PROVIDER: S-EPMC8236421 | biostudies-literature |
REPOSITORIES: biostudies-literature