Hemodynamic monitoring by intracardiac impedance measured by cardiac resynchronization defibrillators: Evaluation in a controlled clinical setting (BIO.Detect HF II study).


ABSTRACT:

Background

In patients with cardiac resynchronization therapy defibrillators (CRT-Ds), intracardiac impedance measured by dedicated CRT-D software may be used to monitor hemodynamic changes. We investigated the relationship of hemodynamic parameters assessed by intracardiac impedance and by echocardiography in a controlled clinical setting.

Methods

The study enrolled 68 patients (mean age, 66 ± 9 years; 74% males) at 12 investigational sites. The patients had an indication for CRT-D implantation, New York Heart Association class II/III symptoms, left ventricular ejection fraction 15%-35%, and a QRS duration ≥150 ms. Two months after a CRT-D implantation, hemodynamic changes were provoked by overdrive pacing. Intracardiac impedance was recorded at rest and at four pacing rates ranging from 10 to 40 beats/min above the resting rate. In parallel, echocardiography measurements were performed. We hypothesized that a mean intra-individual correlation coefficient (rmean) between stroke impedance (difference between end-systolic and end-diastolic intracardiac impedance) measured by CRT-D and the aortic velocity time integral (i.e., stroke volume) determined by echocardiography would be significantly larger than 0.65.

Results

The hypothesis was evaluated in 40 patients with complete data sets. The rmean was 0.797, with a lower confidence interval bound of 0.709. The study hypothesis was met (p = 0.007). A stepwise reduction of stroke impedance and stroke volume was observed with increasing heart rate.

Conclusions

Intracardiac impedance measured by implanted CRT-Ds correlated well with the aortic velocity time integral (stroke volume) determined by echocardiography. The impedance measurements bear potential and are readily available technically, not requiring implantation of additional material beyond standard CRT-D system.

SUBMITTER: Delnoy PHM 

PROVIDER: S-EPMC8263311 | biostudies-literature | 2021 Jul-Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Hemodynamic monitoring by intracardiac impedance measured by cardiac resynchronization defibrillators: Evaluation in a controlled clinical setting (BIO.Detect HF II study).

Delnoy Peter-Paul Henri Marie PHM   Gutleben Klaus-Jürgen KJ   Bruun Niels Eske NE   Maier Sebastian K G SKG   Oswald Hanno H   Stellbrink Christoph C   Johansen Jens Brock JB   Paule Stefan S   Søgaard Peter P  

Indian pacing and electrophysiology journal 20210415 4


<h4>Background</h4>In patients with cardiac resynchronization therapy defibrillators (CRT-Ds), intracardiac impedance measured by dedicated CRT-D software may be used to monitor hemodynamic changes. We investigated the relationship of hemodynamic parameters assessed by intracardiac impedance and by echocardiography in a controlled clinical setting.<h4>Methods</h4>The study enrolled 68 patients (mean age, 66 ± 9 years; 74% males) at 12 investigational sites. The patients had an indication for CRT  ...[more]

Similar Datasets

| S-EPMC10247992 | biostudies-literature
| S-EPMC11545944 | biostudies-literature
| S-EPMC2875184 | biostudies-literature
| S-EPMC11624373 | biostudies-literature
| S-EPMC10023891 | biostudies-literature
| S-EPMC6752801 | biostudies-literature
| S-EPMC7526915 | biostudies-literature
| S-EPMC8174266 | biostudies-literature
| S-EPMC6715540 | biostudies-literature