Performance of five rapid serological tests in mild-diseased subjects using finger prick blood for exposure assessment to SARS-CoV-2.


ABSTRACT:

Objectives

Assess the performance of five SARS-CoV-2 rapid serological tests (RST) using finger prick (FP) blood on-site to evaluate their usability for exposure assessment in population-based seroprevalence studies.

Study design

Since cross-reactivity with common cold human coronaviruses occurs, serological testing includes a risk of false-positive results. Therefore, the selected cohort for RST-validation was based on combined immunoassay (presence of specific antibodies) and RT-qPCR (presence of SARS-CoV-2) data. RST-performance for FP blood and serum was assessed by performing each RST in two groups, namely SARSCoV- 2 positive (n=108) and negative healthcare workers (n=89). Differences in accuracy and positive and negative predictive values (PPV, NPV) were calculated for