Evidence-Based Planning Target Volume Margin Reduction for Modern Lung Stereotactic Ablative Radiation Therapy Using Deformable Registration.


ABSTRACT:

Purpose

Standard planning target volume (PTV) margins for lung stereotactic ablative radiation therapy (SABR) are 5 mm. High-dose-rate volumetric modulated arc therapy delivered using flattening filter-free (FFF) beams with modern immobilization systems may allow for PTV margin reduction. This study assesses whether PTV margins can be reduced from 5 to 3 mm.

Methods

Target intrafractional motions derived from pretreatment and posttreatment cone beam computed tomography (CBCT) scans for 33 patients receiving lung SABR treated with 10XFFF energy and 5-mm PTV margins from 2016 to 2019 were used to calculate the required PTV margin. Deformable registration of the planning CT scan and internal gross tumor volume (IGTV) contour to posttreatment CBCT scans for 36 consecutive patien