Acute safety, efficacy, and advantages of a novel cryoballoon ablation system for pulmonary vein isolation in patients with paroxysmal atrial fibrillation: initial clinical experience.


ABSTRACT:

Aims

Cryoballoon pulmonary vein isolation (PVI) is a safe and effective treatment for atrial fibrillation (AF). Current limitations include incomplete vein occlusion due to balloon rigidity and inconsistent electrogram recording, which impairs identification of isolation. We aimed to evaluate the acute safety and performance of a novel cryoballoon system.

Methods and results

The system includes a steerable sheath, mapping catheter, and a balloon that maintains uniform inflation pressure and size following initiation of ablation. Protocol-directed cryoablation was delivered for 180 s for isolation documented in ≤60 s, otherwise freeze duration was 240 s. Primary endpoints were acute safety and vein isolation. Pulmonary vein isolation was confirmed at ≥30 min post-isolation. D

SUBMITTER: Anic A 

PROVIDER: S-EPMC8350865 | biostudies-literature | 2021 Aug