ABSTRACT: Background
During cardiogenic shock blood circulation is minimal in the human body and does not suffice to survive. The extracorporeal life support system (ECLS) acts as a miniature heart-lung-machine that can be temporarily implanted over major vessels e.g. at the groin of the patient to bridge cardiogenic shock. To perform this procedure in an emergency, a proper training model is desirable. Therefore, a 3-dimensional-printable (3D) material must be found that mimics large vessel needle penetration properties. A suitable test bench for material comparison is desirable.Methods
A test setup was built, which simulated the clinically relevant wall tension in specimens. The principle was derived from an existing standardized needle penetration test. After design, the setup was fabricated by means of 3D printing and mounted onto an universal testing machine. For testing the setup, a 3D printable polymer with low Shore A hardness and porcine aorta were used. The evaluation was made by comparing the curves of the penetration force to the standardized test considering the expected differences.Results
3D printing proved to be suitable for manufacturing the test setup, which finally was able to mimic wall tension as if under blood pressure and penetration angle. The force displacement diagrams showed the expected curves and allowed a conclusion to the mechanical properties of the materials. Although the materials forces deviated between the porcine aorta and the Agilus30 polymer, the graphs showed similar but still characteristic curves.Conclusions
The test bench provided the expected results and was able to show the differences between the two materials. To improve the setup, limitations has been discussed and changes can be implemented without complications.
SUBMITTER: von Steuben T
PROVIDER: S-EPMC8362236 | biostudies-literature |
REPOSITORIES: biostudies-literature