Unknown

Dataset Information

0

Improved precision of noise estimation in CT with a volume-based approach.


ABSTRACT: Assessment of image noise is a relevant issue in computed tomography (CT). Noise is routinely measured by the standard deviation of density values (Hounsfield units, HU) within a circular region of interest (ROI). We explored the effect of a spherical volume of interest (VOI) on noise measurements. Forty-nine chronic obstructive pulmonary disease patients underwent CT with clinical protocol (regular dose [RD], volumetric CT dose index [CTDIvol] 3.04 mGy, 64-slice unit), and ultra-low dose (ULD) protocol (median CTDIvol 0.38 mGy, dual-source unit). Noise was measured in 27 1-cm2 ROIs and 27 0.75-cm3 VOIs inside the trachea. Median true noise was 21 HU (range 17-29) for RD-CT and 33 HU (26-39) for ULD-CT. The VOI approach resulted in a lower mean distance between limits of agreement compared to ROI: 5.9 versus 10.0 HU for RD-CT (-40%); 4.7 versus 9.9 HU for ULD-CT (-53%). Mean systematic bias barely changed: -1.6 versus -0.9HU for RD-CT; 0.0 to 0.4HU for ULD-CT. The average measurement time was 6.8 s (ROI) versus 9.7 (VOI), independent of dose level. For chest CT, measuring noise with a VOI-based instead of a ROI-based approach reduces variability by 40-53%, without a relevant effect on systematic bias and measurement time.

SUBMITTER: Wisselink HJ 

PROVIDER: S-EPMC8429536 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC9612153 | biostudies-literature
| S-EPMC5690921 | biostudies-literature
| S-EPMC4560645 | biostudies-literature
| S-EPMC6657110 | biostudies-literature
| S-EPMC7291325 | biostudies-literature
| S-EPMC7887671 | biostudies-literature
| S-EPMC6371781 | biostudies-literature
| S-EPMC9345340 | biostudies-literature
| S-EPMC6719274 | biostudies-literature
| S-EPMC8587190 | biostudies-literature