Unknown

Dataset Information

0

A gain series method for accurate EMCCD calibration.


ABSTRACT: Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model more accurately estimates the gain response of the camera. Furthermore, we develop a method to compare the results from different calibration approaches. Gain-series calibration outperforms all other methods in this self-consistency test.

SUBMITTER: Ryan DP 

PROVIDER: S-EPMC8443689 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC4634452 | biostudies-literature
| PRJEB24755 | ENA
| S-EPMC8225824 | biostudies-literature
| S-EPMC2988285 | biostudies-other
| S-EPMC4466524 | biostudies-literature
| S-EPMC7648787 | biostudies-literature
| S-EPMC10499875 | biostudies-literature
| S-EPMC5087901 | biostudies-literature
| S-EPMC6519131 | biostudies-literature
| S-EPMC3292602 | biostudies-other