A gain series method for accurate EMCCD calibration.


ABSTRACT: Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model more accurately estimates the gain response of the camera. Furthermore, we develop a method to compare the results from different calibration approaches. Gain-series calibration outperforms all other methods in this self-consistency test.