Different Effects of Aging on Intraocular Pressures Measured by Three Different Tonometers.
Ontology highlight
ABSTRACT: This study aimed to compare intraocular pressures (IOP) using different tonometers, Goldmann applanation (IOPGAT), non-contact (IOPNCT), and rebound (IOPRBT), and to assess the effects of aging and central corneal thickness (CCT) on the measurements. The IOPGAT, IOPNCT, IOPRBT, mean patient age (65.1 ± 16.2 years), and CCT (521.7 ± 39.2 µm) were collected retrospectively from 1054 eyes. The differences among IOPs were compared by the paired t-test. Possible correlations between devices, age, and CCT were assessed by linear regression analyses. The effects of age and CCT on the IOP reading were assessed by mixed-effects regression models. The IOPGAT values were 2.4 and 1.4 mmHg higher than IOPNCT and IOPRBT, respectively; the IOPNCT was 1.0 mmHg lower than IOPRBT (p < 0.0001 for all compari