ABSTRACT: This study aimed to compare intraocular pressures (IOP) using different tonometers, Goldmann applanation (IOPGAT), non-contact (IOPNCT), and rebound (IOPRBT), and to assess the effects of aging and central corneal thickness (CCT) on the measurements. The IOPGAT, IOPNCT, IOPRBT, mean patient age (65.1 ± 16.2 years), and CCT (521.7 ± 39.2 µm) were collected retrospectively from 1054 eyes. The differences among IOPs were compared by the paired t-test. Possible correlations between devices, age, and CCT were assessed by linear regression analyses. The effects of age and CCT on the IOP reading were assessed by mixed-effects regression models. The IOPGAT values were 2.4 and 1.4 mmHg higher than IOPNCT and IOPRBT, respectively; the IOPNCT was 1.0 mmHg lower than IOPRBT (p < 0.0001 for all comparisons). The IOPs measured by each tonometer were highly correlated with each other (r = 0.81-0.90, t = 45.2-65.5). The linear regression analyses showed that age was negatively correlated with IOPNCT (r = -0.12, t = -4.0) and IOPRBT (r = -0.14, t = -4.5) but not IOPGAT (r = 0.00, t = -0.2); the CCT was positively correlated with IOPGAT (r = 0.13, t = 4.3), IOPNCT (r = 0.29, t = 9.8), and IOPRBT (r = 0.22, t = 7.2). The mixed-effect regression models showed significant negative correlations between age and IOPNCT (t = -2.6) and IOPRBT (t = -3.4), no correlation between age and IOPGAT (t = 0.2), and a significant positive correlation between CCT and the tonometers (t = 3.4-7.3). No differences between IOPGAT and IOPRBT were seen at the age of 38.8 years. CCT affects IOPs from all tonometers; age affects IOPNCT and IOPRBT in different degrees. IOPRBT tended to be higher than IOPGAT in young subjects, but this stabilized in middle age and became higher in older subjects.