ABSTRACT: Purpose
This study evaluated the efficacy and ocular surface status of Breath-O Correct, novel orthokeratology (OK) lenses, worn overnight for 3 months. Lens-induced changes in the tear proteome were evaluated.Methods
Thirty-one subjects, aged 19 to 26 years with refractive error from -1.00 to -5.00 D, were randomly assigned 1:1 to the treatment or control group. Refraction, visual acuity, corneal integrity, biomechanics and endothelial health, ocular surface changes, and subjective symptoms were assessed at the baseline, one-month, and three-month visits. The tear proteome was characterized over time using sequential window acquisition of all theoretical ion spectra mass spectrometry.Results
Lenses improved uncorrected visual acuity and reduced spherical powers with similar efficacy to other OK lenses. Significant reductions (P < 0.05) in corneal hysteresis (11.12 ± 1.12 to 10.38 ± 1.36 mm Hg) and corneal resistance factor (11.06 ± 1.32 to 9.90 ± 1.45 mm Hg) were observed in the treatment group after one month of lens wear, whereas other assessed factors remained unchanged. Thirteen and eight differentially expressed proteins were found after one month and three months of lens wear, respectively. Two proteins (proline-rich protein 27 and immunoglobulin V regions) were differentially expressed at both visits.Conclusions
Over a three-month period, Breath-O Correct lenses were overall safe, well tolerated, efficacious in refractive power reduction, and comparable with other OK lenses. Furthermore, their use caused only minor noninflammatory protein expression changes in the tear proteome.Translational relevance
This study investigated the safety of orthokeratology contact lenses on the ocular surface in molecular aspects and standard clinical parameters.