Ontology highlight
ABSTRACT: Purpose
To report the initial safety and efficacy results of a second-generation (44-channel) suprachoroidal retinal prosthesis at 56 weeks after device activation.Methods
Four subjects, with advanced retinitis pigmentosa and bare-light perception only, enrolled in a phase II trial (NCT03406416). A 44-channel electrode array was implanted in a suprachoroidal pocket. Device stability, efficacy, and adverse events were investigated at 12-week intervals.Results
All four subjects were implanted successfully and there were no device-related serious adverse events. Color fundus photography indicated a mild postoperative subretinal hemorrhage in two recipients, which cleared spontaneously within 2 weeks. Optical coherence tomography confirmed device stability and position under the macula. Screen-based localization accuracy was significantly better for all subjects with device on versus device off. Two subjects were significantly better with the device on in a motion discrimination task at 7, 15, and 30°/s and in a spatial discrimination task at 0.033 cycles per degree. All subjects were more accurate with the device on than device off at walking toward a target on a modified door task, localizing and touching tabletop objects, and detecting obstacles in an obstacle avoidance task. A positive effect of the implant on subjects' daily lives was confirmed by an orientation and mobility assessor and subject self-report.Conclusions
These interim study data demonstrate that the suprachoroidal prosthesis is safe and provides significant improvements in functional vision, activities of daily living, and observer-rated quality of life.Translational relevance
A suprachoroidal prosthesis can provide clinically useful artificial vision while maintaining a safe surgical profile.
SUBMITTER: Petoe MA
PROVIDER: S-EPMC8479573 | biostudies-literature | 2021 Aug
REPOSITORIES: biostudies-literature
Petoe Matthew A MA Titchener Samuel A SA Kolic Maria M Kentler William G WG Abbott Carla J CJ Nayagam David A X DAX Baglin Elizabeth K EK Kvansakul Jessica J Barnes Nick N Walker Janine G JG Epp Stephanie B SB Young Kiera A KA Ayton Lauren N LN Luu Chi D CD Allen Penelope J PJ
Translational vision science & technology 20210801 10
<h4>Purpose</h4>To report the initial safety and efficacy results of a second-generation (44-channel) suprachoroidal retinal prosthesis at 56 weeks after device activation.<h4>Methods</h4>Four subjects, with advanced retinitis pigmentosa and bare-light perception only, enrolled in a phase II trial (NCT03406416). A 44-channel electrode array was implanted in a suprachoroidal pocket. Device stability, efficacy, and adverse events were investigated at 12-week intervals.<h4>Results</h4>All four subj ...[more]