ABSTRACT: Objectives
Computed tomography (CT) allows reproducible assessment of left ventricular (LV) function, left ventricular outflow tract area (LVOTarea) and aortic valve area (AVA). We evaluated the influence of image reconstruction parameters on these measurements.Methods
We analyzed 45 contrast-enhanced, retrospectively ECG-gated CT datasets acquired on a third-generation dual source system. A standard filtered-back-projection data set (20 cardiac phases (5% steps, 0-95%), 0.6-mm-slice thickness, 512 × 512 matrix) and eight reconstructions with modified slice thickness (1-8 mm), number of cardiac phases (5, 10), matrix size (256×256) and an iterative reconstruction (IR) algorithm were obtained. LV parameters (ejection fraction (EF), stroke volume (SV), end-diastolic (EDV), end-systolic volumes (ESV)), LVOTarea and AVA were assessed.Results
Differences in LV parameters, LVOTarea and AVA, were only minimal between standard reconstructions and those with modified matrix size, IR algorithm and ≤2 mm slice thickness, while reconstructions with 8-mm slice thickness significantly overestimated SV (p < 0.001) and EDV (p = 0.016). AVA planimetry in reconstructions with ≥5 mm slice thickness was not feasible in 56% of patients. A decrease in the number of reconstructed phases (10 or 5) underestimated EF, SV, EDV, LVOTarea and AVA and overestimated ESV.Conclusions
Modifications of reconstruction parameters (except a slice thickness ≤2 mm) have only a marginal effect on LV, LVOTarea and AVA assessment. However, a reduced number of reconstructions per cardiac cycle may significantly influence measurements.Advances in knowledge
Substantial modifications in number of reconstructions per cardiac cycle significantly affect the assessment of LV function, LVOTarea and AVA also in modern CT scanners.