Evaluation of a Digital Handheld Hydrogen Breath Monitor to Diagnose Lactose Malabsorption: Interventional Crossover Study.


ABSTRACT:

Background

Lactose malabsorption is a common condition that affects a broad segment of the population. Clinical diagnosis based on symptom recall can be unreliable and conventional testing can be inconvenient, requiring expensive laboratory-based equipment and conduction of the testing in a clinical setting.

Objective

The aim of this study is to assess the performance of a digital handheld hydrogen breath monitor (GIMate) in diagnosing lactose malabsorption compared to a US Food and Drug Administration (FDA)-cleared device (H2 Check) for the same indication.

Methods

An interventional crossover study was performed in adult participants with a prior confirmed diagnosis of lactose malabsorption or a suspected history of lactose intolerance.

Results

A total of 31 participants (mean age 33.9 years) were enrolled in the study. There was 100% positive percent agreement and 100% negative percent agreement between the GIMate monitor and the H2 Check. Correlation between gastrointestinal symptoms and hydrogen values was positive at 0.82 (P<.001).

Conclusions

The digital handheld GIMate breath monitor achieved equivalent diagnostic performance to that of an FDA-cleared device in the diagnosis of lactose malabsorption.

Trial registration

ClinicalTrials.gov NCT04754724; https://clinicaltrials.gov/ct2/show/NCT04754724.

SUBMITTER: Mathews SC 

PROVIDER: S-EPMC8561400 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7325051 | biostudies-literature
| S-EPMC7812489 | biostudies-literature
| S-EPMC5268498 | biostudies-literature
| S-EPMC10132719 | biostudies-literature
| S-EPMC5597491 | biostudies-other
| S-EPMC6839734 | biostudies-literature
| S-EPMC3672687 | biostudies-literature
| S-EPMC4311178 | biostudies-literature
| S-EPMC6107192 | biostudies-literature
| S-EPMC6950325 | biostudies-literature