Decision Threshold for Kryptor sFlt-1/PlGF Ratio in Women With Suspected Preeclampsia: Retrospective Study in a Routine Clinical Setting.


ABSTRACT: Background The objective was to evaluate predictive performance and optimal decision threshold of the Kryptor soluble fms-like tyrosine kinase-1 (sFlt-1)/placental growth factor (PlGF) ratio when implemented for routine management of women presenting with symptoms of preeclampsia. Methods and Results Observational retrospective study of a cohort of 501 women with suspected preeclampsia after 20 weeks of gestation. Women referred to maternity ward for observation of preeclampsia had an sFlt-1/PlGF ratio test included in routine diagnostic workup. Maternal and offspring characteristic data included maternal risk factors, outcomes, delivery mode, and indication for suspected preeclampsia. Biochemical measurements to determine sFlt-1/PlGF ratio were performed using the BRAHMS/Kryptor sFlt-1/PlGF ratio immunoassays. Results were analyzed by area under receiver-operating characteristic curve. Preeclampsia occurred in 150 of 501 (30%) of symptomatic women with an sFlt-1/PlGF ratio determined before the time of diagnosis. Area under receiver-operating characteristic curve for diagnosis of early-onset preeclampsia within 1 and 4 weeks was 0.98 (95% CI, 0.96-1.00) and 0.95 (95% CI, 0.92-0.98), respectively. For late-onset preeclampsia, predictive performance within 1 and 4 weeks was lower: 0.90 (95% CI, 0.85-0.94) and 0.85 (95% CI, 0.80-0.90), respectively. The optimal single sFlt-1/PlGF ratio threshold for all preeclampsia and late-onset preeclampsia within 1 and 4 weeks was 66. The negative and positive predictive values for ruling out and ruling in developing preeclampsia within 1 week were 96% and 70%, respectively. Conclusions The Kryptor sFlt-1/PlGF ratio is a useful clinical tool ruling out and in preeclampsia within 1 week. Prediction within 4 weeks is superior for early-onset preeclampsia. A single decision threshold of 66 is indicated for use in clinical routine.

SUBMITTER: Andersen LLT 

PROVIDER: S-EPMC8649230 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC4887119 | biostudies-literature
| S-EPMC6710794 | biostudies-literature
| S-EPMC6080558 | biostudies-literature
| S-EPMC8255209 | biostudies-literature
| S-EPMC9866466 | biostudies-literature
| S-EPMC8255211 | biostudies-literature
| S-EPMC5215695 | biostudies-literature
| S-EPMC9727903 | biostudies-literature
| S-EPMC8117392 | biostudies-literature
| S-EPMC3864477 | biostudies-other