Proof-of-concept and concurrent validity of a prototype headset to assess peak oxygen uptake without a face mask.


ABSTRACT:

Objective

Portable gas exchange instruments allow the assessment of peak oxygen uptake (V̇O2peak) but are often bulky, expensive and require wearing a face mask thereby limiting their routine application. A newly developed miniaturized headset (VitaScale, Nuremberg, Germany) may overcome these barriers and allow measuring V̇O2peak without applying a face mask. Here we aimed (i) to disclose the technical setup of a headset incorporating a gas and volume sensor to measure volume flow and expired oxygen concentration and (ii) to assess the concurrent criterion-validity of the headset to measure V̇O2peak in 44 individuals exercising on a stationary cycle ergometer in consideration of the test-retest reliability of the criterion measure.

Results

The coefficient of variation (CV%) while measuring V̇O2peak during incremental cycling with the headset was 6.8%. The CV% for reliability of the criterion measure was 4.0% for V̇O2peak. Based on the present data, the headset might offer a new technology for V̇O2peak measurement due to its low-cost and mask-free design.

SUBMITTER: Duking P 

PROVIDER: S-EPMC8742938 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7783007 | biostudies-literature
| S-EPMC3604641 | biostudies-literature
| S-EPMC7928787 | biostudies-literature
| S-EPMC7273233 | biostudies-literature
| S-EPMC9355742 | biostudies-literature
| S-EPMC7961526 | biostudies-literature
| S-EPMC10700517 | biostudies-literature
| S-EPMC3157615 | biostudies-literature
| S-EPMC7995042 | biostudies-literature
| S-EPMC7313704 | biostudies-literature