Unknown

Dataset Information

0

Practical implementation of remote continuous glucose monitoring in hospitalized patients with diabetes.


ABSTRACT:

Disclaimer

In an effort to expedite the publication of articles, AJHP is posting manuscripts online as soon as possible after acceptance. Accepted manuscripts have been peer-reviewed and copyedited, but are posted online before technical formatting and author proofing. These manuscripts are not the final version of record and will be replaced with the final article (formatted per AJHP style and proofed by the authors) at a later time.

Purpose

Inpatient diabetes management involves frequent assessment of glucose levels for treatment decisions. Here we describe a program for inpatient real-time continuous glucose monitoring (rtCGM) at a community hospital and the accuracy of rtCGM-based glucose estimates.

Methods

Adult inpatients with preexisting diabetes managed with intensive insulin therapy and a diagnosis of coronavirus disease 2019 (COVID-19) were monitored via rtCGM for safety. An rtCGM system transmitted glucose concentration and trending information at 5-minute intervals to nearby smartphones, which relayed the data to a centralized monitoring station. Hypoglycemia alerts were triggered by rtCGM values of ≤85 mg/dL, but rtCGM data were otherwise not used in management decisions; insulin dosing adjustments were based on blood glucose values measured via blood sampling. Accuracy was evaluated retrospectively by comparing rtCGM values to contemporaneous point-of-care (POC) blood glucose values.

Results

A total of 238 pairs of rtCGM and POC data points from 10 patients showed an overall mean absolute relative difference (MARD) of 10.3%. Clarke error grid analysis showed 99.2% of points in the clinically acceptable range, and surveillance error grid analysis showed 89.1% of points in the lowest risk category. It was determined that for 25% of the rtCGM values, discordances in rtCGM and POC values would likely have resulted in different insulin doses. Insulin dose recommendations based on rtCGM values differed by 1 to 3 units from POC-based recommendations.

Conclusion

rtCGM for inpatient diabetes monitoring is feasible. Evaluation of individual rtCGM-POC paired values suggested that using rtCGM data for management decisions poses minimal risks to patients. Further studies to establish the safety and cost implications of using rtCGM data for inpatient diabetes management decisions are warranted.

SUBMITTER: Baker M 

PROVIDER: S-EPMC8767852 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC8323182 | biostudies-literature
| S-EPMC5601025 | biostudies-literature
| S-EPMC2769750 | biostudies-literature
| S-EPMC5770096 | biostudies-literature
| 71713 | ecrin-mdr-crc
| S-EPMC3064030 | biostudies-literature