Feasibility and Accuracy of Thoracolumbar Pedicle Screw Placement Using an Augmented Reality Head Mounted Device.


ABSTRACT:

Background

To investigate the accuracy of augmented reality (AR) navigation using the Magic Leap head mounted device (HMD), pedicle screws were minimally invasively placed in four spine phantoms.

Methods

AR navigation provided by a combination of a conventional navigation system integrated with the Magic Leap head mounted device (AR-HMD) was used. Forty-eight screws were planned and inserted into Th11-L4 of the phantoms using the AR-HMD and navigated instruments. Postprocedural CT scans were used to grade the technical (deviation from the plan) and clinical (Gertzbein grade) accuracy of the screws. The time for each screw placement was recorded.

Results

The mean deviation between navigation plan and screw position was 1.9 ± 0.7 mm (1.9 [0.3-4.1] mm) at the entry point and 1.4 ± 0.8 mm (1.2 [0.1-3.9] mm) at the screw tip. The angular deviation was 3.0 ± 1.4° (2.7 [0.4-6.2]°) and the mean time for screw placement was 130 ± 55 s (108 [58-437] s). The clinical accuracy was 94% according to the Gertzbein grading scale.

Conclusion

The combination of an AR-HMD with a conventional navigation system for accurate minimally invasive screw placement is feasible and can exploit the benefits of AR in the perspective of the surgeon with the reliability of a conventional navigation system.

SUBMITTER: Frisk H 

PROVIDER: S-EPMC8779462 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC3153609 | biostudies-literature
| S-EPMC8819976 | biostudies-literature
| S-EPMC4824416 | biostudies-literature
| S-EPMC10464480 | biostudies-literature
| S-EPMC5502290 | biostudies-other
| S-EPMC7668595 | biostudies-literature
| S-EPMC9134485 | biostudies-literature
| S-EPMC9193845 | biostudies-literature
| S-EPMC10267819 | biostudies-literature