The Use of an Acylhydrazone-Based Metal-Organic Framework in Solid-Contact Potassium-Selective Electrode for Water Analysis.


ABSTRACT: A solid-contact ion-selective electrode was developed for detecting potassium in environmental water. Two versions of a stable cadmium acylhydrazone-based metal organic framework, i.e., JUK-13 and JUK-13_H2O, were used for the construction of the mediation layer. The potentiometric and electrochemical characterizations of the proposed electrodes were carried out. The implementation of the JUK-13_H2O interlayer is shown to improve the potentiometric response and stability of measured potential. The electrode exhibits a good Nernstian slope (56.30 mV/decade) in the concentration range from 10-5 to 10-1 mol L-1 with a detection limit of 2.1 µmol L-1. The long-term potential stability shows a small drift of 0.32 mV h-1 over 67 h. The electrode displays a good selectivity comparable to ion-selective electrodes with the same membrane. The K-JUK-13_H2O-ISE was successfully applied for the determination of potassium in three certified reference materials of environmental water with great precision (RSD < 3.00%) and accuracy (RE < 3.00%).

SUBMITTER: Koscielniak P 

PROVIDER: S-EPMC8781722 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

The Use of an Acylhydrazone-Based Metal-Organic Framework in Solid-Contact Potassium-Selective Electrode for Water Analysis.

Kościelniak Paweł P   Dębosz Marek M   Wieczorek Marcin M   Migdalski Jan J   Szufla Monika M   Matoga Dariusz D   Kochana Jolanta J  

Materials (Basel, Switzerland) 20220113 2


A solid-contact ion-selective electrode was developed for detecting potassium in environmental water. Two versions of a stable cadmium acylhydrazone-based metal organic framework, i.e., JUK-13 and JUK-13_H2O, were used for the construction of the mediation layer. The potentiometric and electrochemical characterizations of the proposed electrodes were carried out. The implementation of the JUK-13_H2O interlayer is shown to improve the potentiometric response and stability of measured potential. T  ...[more]

Similar Datasets

| S-EPMC5879484 | biostudies-literature
| S-EPMC6641300 | biostudies-literature
| S-EPMC4451844 | biostudies-literature
| S-EPMC11536185 | biostudies-literature
| S-EPMC5449066 | biostudies-other
| S-EPMC8651668 | biostudies-literature
| S-EPMC9175116 | biostudies-literature
| S-EPMC5713653 | biostudies-literature
| S-EPMC7467549 | biostudies-literature
| S-EPMC10976635 | biostudies-literature